Ambulatory Blood Pressure Monitor 90217 Service Manual
Time of Day
A real-time clock provides the time of day, which appears between measurement cycles and can
be programmed in either a 12- or 24-hour mode.
Cuff Pressure
This pressure can be shown while the monitor is taking a measurement. If cuff pressure is not
selected, "----" appears. Cuff pressure can be enabled or disabled in software.
Event Code
Event codes are 4-digit LCD messages that begin with "EC" or, in some cases, show specific
codes such as "LLL" for a low battery. If enabled, a beep sounds during an event code and the
code is displayed on the monitor. Refer Troubleshooting on page 5-1 for event code information.
Count Down Sequence
Whenever the monitor begins a measurement cycle, it turns its tone ON (if beep is enabled) and
counts down from 5555 to 1111 on the display.
IR Communications
During IR communications, the monitor first displays 9999 on the LCD to indicate that it has
detected a cable connect and has gone into the communication mode. The monitor periodically
attempts to contact a modem. The left two digits on the monitor display indicate the steps in the
communication process. For more information, refer to Base Station via Modem on page 2-3.

Programming Options

These monitor options are user programmed:
Day or Night Modes
Day mode = beeper ON, inflation time approximately 15 seconds
Night mode = beeper OFF, inflation time approximately 20 seconds
Measurement Intervals and Periods
Measurement intervals are selectable from 6 to 120 minutes (in one minute increments) for each
period. It is also possible to not have readings taken during a specified period. Up to 12 periods
may be defined.
Clinical Verification Mode
When set to this mode, the monitor is forced to bleed to 40 mmHg or to one step below diastole
(whichever is lower) for each reading.