Disassembly Procedures - Spacelabs 90217 Service Manual

Ambulatory blood pressure monitor
Hide thumbs Also See for 90217:
Table of Contents

Advertisement

Ambulatory Blood Pressure Monitor 90217 Service Manual

Disassembly Procedures

Caution
Observe precautions for handling electrostatic-sensitive devices!
Note:
Never touch electrostatic-sensitive electronic components without following proper
antistatic procedures, including the use of an ESD wrist band and mat. An
electrostatic discharge from your fingers can permanently damage electronic
components.
All static-sensitive electronic components are packaged in static-shielding bags.
Retain this bag for repackaging the component should you need to store it or
return it to Spacelabs Medical for any reason.
To open the monitor and access its internal components:
1 Remove the Communications label from the back of the unit (refer to Parts on page 6-1 of this
manual for replacement part number) and remove the two screws that are exposed (see
below).
2 Remove the identify label on top of the unit (see illustration below) and remove the two exposed
screws to separate the top and bottom case pieces.
To separate the ECB assembly from the bottom case half:
1 Remove the front panel label from the unit (refer to Parts on page 6-1 of this manual for
replacement part number) and expose the two screws holding the ECB assembly.
2 Remove these two screws (see previous illustration) and separate the ECB assembly from the
lower half of the case.
4-15