Maintenance; Cleaning - Spacelabs 90217 Service Manual

Ambulatory blood pressure monitor
Hide thumbs Also See for 90217:
Table of Contents

Advertisement

Contents

Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Calibration Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operational Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Manifold Kit - P/N 050-0110-xx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Periodic maintenance consists of cleaning the unit, replacing or recharging the batteries, testing
the unit for accurate operation, and calibrating when necessary.
Cleaning
Use a soft, damp cloth and mild detergent mixed with water to wipe the exterior of the monitor.
Clean the carrying pouch and air hose with isopropyl alcohol.
The cuff wrap may be sterilized only with ethylene oxide (ETO) sterilization methods using
standard hospital procedures. Use standard aeration techniques after sterilization. Small soiled or
stained areas may be cleaned by gentle scrubbing with a sponge or cloth soaked in a mild soap
and water solution.
The cuff wrap with the air bladder removed is machine washable on "delicate" cycle only. Do not
wash with bed linens, gowns, or in large commercial-type washers.
To remove the bladder for cleaning, refer to the figures below and follow these steps:
1 Using your fingers only, fold or roll up the bladder inside the cuff. Do not use pencils, pens, or
other hard objects as damage to the bladder could easily occur.
2 Remove the bladder through the hose exit opening. Once the bladder has been removed, be
sure to attach the hook and loop surfaces on the cuff before washing.
3 After washing and drying the cuff, reinstall the bladder in the reverse order of its removal. Make
certain that all folds in the bladder are removed prior to inserting it back inside the cuff.
Note:
The bladder may be installed with the hose exiting the second cuff opening. However,
the bladder must be positioned with its long side toward the center of the cuff.

Maintenance

4-1