Base Station Report Event Codes - Spacelabs 90217 Service Manual

Ambulatory blood pressure monitor
Hide thumbs Also See for 90217:
Table of Contents

Advertisement

Ambulatory Blood Pressure Monitor 90217 Service Manual

Base Station Report Event Codes

The following list contains the extended event codes, which may appear in a blood pressure
report. The extended event code digit appears in the first (tens) digit position (for example, 11).
The list is grouped according to the monitor event code (if applicable), which would be displayed at
the time of the event. The codes are printed as a numeric value in the systolic column with all other
columns printing zero.
Monitor displays -- EC00
Base Station report prints:
10
20
30
40
50
Monitor displays -- EC01
Base Station report prints:
11
21
Monitor displays -- EC02
Base station report prints:
12
22
32
42
52
62
72
82
Monitor displays -- EC03
Base station report prints:
03
Measurement aborted as the result of excess movement artifact. Frequent "10"
messages may indicate an air leak.
A) A very large number of movement artifacts.
B) Heart rate arrhythmia.
A) Movement artifact at mean arterial pressure.
B) Heart rate arrhythmia.
A) Movement artifact at systole.
B) Heart rate arrhythmia.
A) Movement artifact at diastole.
B) Heart rate arrhythmia.
Did not pump above the mean arterial pressure.
Did not pump above the systolic pressure.
Did not reach initial cuff pressure. The cuff may have been improperly applied or
there may be an air leak.
Overpressure.
Overpressure.
No cuff attached.
Kinked hose.
Cuff applied too loosely.
Kinked hose.
Kinked hose.
Patient canceled measurement by pressing monitor's START/STOP button.
5-3