RECOMMENDED MAINTENANCE AND INSPECTION

Maintenance Inspection of All
Components (Receipt of shipment
• Make sure all parts/components are included (Please see
page 8 - "Unpacking Your Bed".
• Check all bed components for obvious damage.
• Inspect the Power Supply Cord for any cuts and/or damage.
• Check to see all actuator cords are routed and connected
properly to the control box.
Mechanical Inspection of Assemblies
(6 months)
• Inspect all welds on the sleeping surface, frame, and base
assemblies for stress fractures.
• Inspect all fasteners for wear and looseness.
• IMPORTANT: Lubricate all pivot points, actuator clevis pins,
and control arm clevis pins as needed. White Lithium Grease
is recommended.
Mechanical Inspection of Casters
& Pedal Locking Mechanism
(3 months)
• Check the pedal locking mechanism to make sure it engages
and disengages properly.
• Check the bottom foot pads on the foot end casters, making
sure they are clean of debris and undamaged. Replace if needed.
• Check all casters to ensure that they roll properly and
are unobstructed.
The warranties contained herein contain all the representations and warranties with respect to the subject matter
of this document, and supersede all prior negotiations, agreements and understandings with respect thereto.
The recipient of this document hereby acknowledges and represents that it has not relied on any representation,
assertion, guarantee, warranty, collateral contract or other assurance, except those set out in this document.
Basic American Medical Products, a division of GF Health Products, Inc.
)
WARRANTY
15 Years on Frame
5 Years on motors and control Box
(longest motor warranty in the industry)
3 Years on Other Electronics
Electrical Inspection of Control Box
(6 months)
• Check the power cord connected to the control box for
any chafing, cuts or wear.
• Make sure all attaching hardware is securely tightened.
• Check all electrical connections for any wear or fractures.
• Check your 9V battery (see page 19). Replace if needed.
Electrical Inspection of Pendant
and Staff Control (6 months)
• Check the pendant cord for any chafing, cuts or wear.
• Check all pendant functions . . .
✓ Head raises and lowers properly
✓ Foot raises and lowers properly
✓ Entire bed raises and lowers properly.
• Check to make sure each button and associated function
work properly (i.e. the head section raises when the Head
Up button is activated).
Electrical Inspection of Actuators
(3 months)
• Check the actuator cords for any chafing, cuts or wear.
• Check the range of movement on all motors to ensure they
do not bind in the Full Up and Full Down positions.
• Make sure the bed is sitting horizontally (parallel to the
floor) when in its full down position. If not, please see
pages 24 and 25.