Troubleshooting - Staff Control; Removing And Replacing The Staff Control Assembly - Basic American Zenith 7000 Instructions For Use And Service Manual

Table of Contents

Advertisement

REMOVING AND REPLACING THE
STAFF CONTROL ASSEMBLY
1. UNPLUG THE POWER CORD FROM THE
WALL OUTLET.
2. Unscrew the 4 Phillips Screws (#1) that attach the
Shroud (cable cover) to the footboard. Set aside.
3. At the foot end of the bed, unscrew the retaining
caps on the male staff control cable (#3) and the
Grey T-Cable (#4) and unplug the male staff
cable & female staff cable (#5) from the Black
Control Cable (#6) and Grey T-Cable.
4. Unscrew the two remaining Phillips Screws (#2)
from the inside of the footboard that attach the
Staff Control Assembly (#7) bevel to the board.
Set aside screws for later re-assembly.
5. Remove the OLD Staff Control Assembly and
fish the old cables out through the large hole.
6. Feed the two cables of the NEW Staff Control
Assembly from the outside through the large
hole and insert the bevel into the cutout on
the outside of the footboard.
7. Secure the Staff Control by re-inserting
the two Phillips Screws (#2) you
removed in Step 4.
8. Re-position the Shroud
on the inside of the
footboard and
secure with the
four Phillips
Screws (#1) you
removed in Step 2.
9. Plug the Male Staff Control
Cable (#3) into the Black Control
Cable (#6), making sure the phone jack plug
is seated correctly into the Black Control Cable.
Screw on the retaining cap to secure.
10. Plug the Female Staff Control Cable (#5) into
the Grey T-Cable (#4), making sure the T-Cable
plug is seated firmly into the Staff Control's
receptacle. Screw on the retaining cap.
11. Run a Cable Tie through the large end hole on
the right-hand footboard mounting tube.
12. Loop the longer male staff control cable and
gather both staff control cables together. Run
the Cable Tie around both cables and pull tight
to secure. Cut off any excess Cable Tie.
Basic American Medical Products, a division of GF Health Products, Inc.

TROUBLESHOOTING - STAFF CONTROL

(#2)
Staff Control
Bevel
Screws
Male Staff
Control Cable
www.grahamfield.com
ZENITH SERIES
(#7)
Staff Control
Assembly with
bevel, panel &
cables attached
(#1)
Shroud
Screws
IMPORTANT: Run a Cable
Tie through the end hole
in the mounting tube,
loop the longer male staff
control cable, run the tie
around both staff control
cables, and pull tie tight
to secure. Cut off excess
cable tie.
Black
Control Cable
999-0843-190A Service Manual
Footboard
Female Staff
Control Cable
T-Cable
CABLE
CONNECTION
CLOSEUP
23