Summary of Contents for Basic American Zenith 7000
Page 1
To avoid personal injury or damage to bed, please read all sections pertaining to your bed model before use. ETL/UL/CSA APPROVED UL 60601-1 IEC 60601-2-38 CSA 22.2 601.1 Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com Zenith SERIES Service Manual 999-0843-190A...
Page 2
GF Health Products, Inc. 2935A Northwest Parkway Atlanta, GA 30360 www.grahamfield.com To order a Zenith 7000 Bed please contact a Graham Field Sales Representative at 1-800-554-9215 Important notIce The information contained in this document is subject to change without notice.
Page 3
Service Part Diagrams ....29-32 label symbol definitions Consult Safe Accompanying Working Documents Load Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 4
Leave it off when unattended to reduce the risk of injury. If patient/ while the transmission is in progress. resident is ambulatory, leave at chair height. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 5
AVENT (TETE) DU LIT. LES COTES DE LAS BED CAN RESULT IN DAMAGE TENTE OXYGENE NE DOIVENT PAS SE TO PROPERTY OR PERSONAL PROLONGER PLUS DAS QUE LA SOMMIER INJURY. DU LIT. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 6
We will help you determine which bed/bed part the safety of the patient/resident. is best for your patient’s/resident’s particular needs and help you with your compliance issues. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 7
Internal Battery Backup Replacement . . . 9V Battery (Note: Ilcon External Battery Pack and Charger can be purchased separately as an accessory - see page 19.) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 8
WIRE ALONG WITH PENDANT, PENDANT CABLE & HOLSTER. PLEASE CUT AND DISCARD CABLE TIE (#13) WHEN YOU UN-PACK YOUR BED. ALSO DISCARD TOP AND BOTTOM CARTONS Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 9
NOTE: The first hole at the foot end is POSITIONS WITH reserved for attaching the Optional 4” 76/80” RETAINER & WALLSAVER POSITIONS pan extension on an 80” bed. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 10
PLUGGING IN THE OPTIONAL FOOTBOARD STAFF CONTROL STEP 1 - ATTACHING THE FOOTBOARD: You can order an optional Staff Control for the Zenith 7000 which includes a Staff Control Assembly and Shroud Cover pre-installed to the Footboard at the manufacturing factory.
Page 11
This gap must be small enough to prevent of an improperly fitted mattress could result in injury resident/patient from getting his/her head or neck or death. caught in this location. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 12
76” bed, position the wallsaver ends in the MIDDLE slots on the caster bases. For 80” beds position wallsaver in slots closest to the HEAD. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 13
ZENITH SERIES BED OPERATIONS - PATIENT PENDANT/HAND CONTROLLER The hand controller comes standard on the Zenith 7000 series and can be plugged into either side of the bed. The plug features a lock-on cap. A pendant holster for attaching to optional assist devices is also provided.
Page 14
NEVER operate the bed if a Power Cord or Plug is damaged or not working properly. Contact qualified Service Personnel for examination and repair. ALWAYS UNPLUG THE POWER CORD WHEN PERFORMING ANY MAINTENANCE ON THE BED. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 15
UNPLUG FOOT END CABLES TO PLUG-IN YOUR STAFF CONTROL ∑ º Control Cable plugs directly into staff cable Pendant T-Cable plugs directly into staff cable (PLEASE SEE PAGE 10) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 16
ARE CLEAR OF THE SIDES OR UNDERNEATH THE IS ALWAYS IN THE FULL, VERTICAL, LOCKED PIVOTING ASSEMBLY WHEN PIVOTING THE POSITION WHEN RESIDENT/PATIENT IS LEFT ASSIST BAR UP OR DOWN. UNATTENDED. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 17
Make sure there are no objects in the way that may hinder the full movement of the rail. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 18
Trapeze frame tubes as shown. Support Adaptor.) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 19
Plugs directly into “optional” ILCOCHARGE Backup Battery & Battery Pack purchased separately. To plug in battery charger, first remove the protective cap with small screwdriver. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 20
Plug Pendant directly into cable and use new cable (See next page Control Cable to test for installation). • NOTHING WORKS - Replace Control Box. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 21
Loop the longest of the staff control cables and tie both together to the tube. Pull tight to secure. Now gently pull the Control Cable’s tie tight. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 22
See page 28 for order information. PENDANT/HAND CONTROLLER • NOTHING WORKS = Replace the Hand Controller. PLEASE USE ORDER KIT See page 28 for order information. # 999-0808-112. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 23
Run CABLE Black the Cable Tie around both cables and pull tight CONNECTION Control Cable to secure. Cut off any excess Cable Tie. CLOSEUP Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 24
Test to see if Hi/Lo Motors need to be Replace the bad Hi/Lo Motor. See re-aligned (See Step F above) page 27 for order information. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 25
RE-ALIGNMENT OF HI/LO MOTORS The Zenith 7000 Hi/Lo Motors are aligned and synchronized to work smoothly in tandem with one another. In the event of a power surge or power outage, or if you are replacing an existing motor, it may be necessary to “re-align” your Hi/Lo motors.
Page 26
The recipient of this document hereby acknowledges and represents that it has not relied on any representation, assertion, guarantee, warranty, collateral contract or other assurance, except those set out in this document. Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com...
Page 27
Installation Instruction Sheet controL caBLe 999-0844-200SP Black Control Cable 1 Cable control cable Hardware pack - IncLUDeS . . . Standard Cable Ties Installation Instruction Sheet Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 28
5/8” round plugs pack (both sides of foot deck) 12/pkg 100-4738-005PK 3/8” round plugs (on main frame rails - foot end) 12/pkg FOR GENERAL CAP/PLUG LOCATION DIAGRAMS, PLEASE SEE PAGES 31 & 32 Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 29
PLEASE HAVE YOUR BED’S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY. (SEE PAGE 5 FOR LOCATION OF SERIAL NUMBER ID LABEL) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 30
PLEASE HAVE YOUR BED’S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY. (SEE PAGE 5 FOR LOCATION OF SERIAL NUMBER ID LABEL) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 31
(PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOOSE ANY CAPS/PLUGS ON YOUR BED - SEE BOTTOM OF PAGE 28 FOR ORDER NUMBERS) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...
Page 32
(PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOOSE ANY CAPS/PLUGS ON YOUR BED - SEE BOTTOM OF PAGE 28 FOR ORDER NUMBERS) Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0843-190A Service Manual...