2.1.2 Alternate BP Field Settings
configured to Chinese)
Mean Arterial Pressure or blood pressure in kPa can be displayed by selecting it on the Advanced
Configuration menu. MAP or kPa is displayed in the upper left corner of the screen and is included on
the printout.
You can display MAP (Mean Arterial Pressure) or
blood pressure in kPa (kiloPascals) in the upper
left corner of the screen, above the ECG
waveform. This alternate BP field display can be
set by using the Advanced Configuration menu.
To set the alternate BP field display , enter the
Advanced Configuration menu:
• Press the
Date/Time
• Press the
Trend
• Use the left
Select
"Alternate BP Field:".
• Press the left
Set
or "Nothing."
• Press the
Trend
Advanced Configuration.
MAP - If MAP is displayed on the screen, it will
also appear in the trend data and in the current
readings of a waveform printout.
MAP Alarms - When either of the MAP limits
are exceeded, an audible alarm sounds and the
affected measurement flashes. If the Print On
Alarm configuration is set to Yes, a MAP alarm
will trigger an automatic print. You can silence
the MAP alarm, and all alarms, by pressing the
large blue
button at the right side of the
Silence
instrument. This will silence all alarms for 60, 90,
or 120 seconds, depending on the setting selected
in Advanced Configuration. However, any
measurement still outside the set limits will flash.
When alarms are silenced, they do not trigger a
Print On Alarm.
To suspend the MAP alarm, press the
button.
button.
button to highlight
to choose "MAP", "kPa",
button again to exit
BP
(only available if language setting
button so the red LED in the button
ALARMS Off
lights. A suspended alarm will still flash if it goes
outside the range of the limits, but it will not
sound the audible alarm. When an alarm is
suspended, it will not trigger a Print On Alarm.
Note: MAP is calculated mathematically from the
Systolic and Diastolic pressures; it is not
measured directly.
Setting alarm limits – If MAP is displayed, you
can set the alarm limits, using the
buttons on the left side of the monitor to follow
these steps:
• Press the left Select button to cycle
through MAP
, Respiration
LO
or 623xx) , then (model 623xx) CO
.
LO
• Each push of Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and
indicating which limit is being changed.
• Press the Set button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
• Press the Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
or
indicators are lit. The instrument is
LO
now in its normal measurement mode. (If
you do not press any button for 10 seconds,
the instrument will automatically revert to
its normal measurement mode).
10
and
Select
, Heart Rate
and
HI
LO
and
(models 622xx
HI
LO
2
or
will flash,
HI
LO
Set
and
HI
and
HI
HI