Welch Allyn Atlas Operator's Manual page 90

Hide thumbs Also See for Atlas:
Table of Contents

Advertisement

PROBLEM:
Blood pressure automatic timing stops.
RECOMMENDED ACTIONS:
Safety mechanisms automatically stop the timed blood pressure measurements when the Atlas Monitor
repeatedly fails to make a measurement within 3 minutes
Determine the reason that the measurements are failing and restart the timed cycle. Press
to manually start a measurement; Auto timing will resume.
Start/Cancel
E.1.2 CO
2
PROBLEM:
Pump does not start.
RECOMMENDED ACTIONS:
Watertrap not fully inserted.
CO
not selected as Second trace selection in Advanced Configuration menu.
2
PROBLEM:
Display does not start - waveform, Respiration Rate, CO
RECOMMENDED ACTIONS:
Watertrap not fully inserted.
CO
not selected as Second trace selection in Advanced Configuration menu.
2
PROBLEM:
Long delay in starting/stabilizing waveform display or numeric displays.
RECOMMENDED ACTIONS:
CO
function is specified to begin display within 10 seconds, but not to reach full specification for up to 5
2
minutes.
Restriction in sampling tubing.
Poor placement of nasal cannula in patient's nose.
Patient breathing orally.
Patient breathing very shallowly.
PROBLEM:
CO
readings inaccurate.
2
RECOMMENDED ACTIONS:
Pressure or suction being applied to sampling tubing – check for flow interference from ventilation,
oxygen, anesthetic supplies.
High concentration of oxygen or nitrous oxide requires compensation calculation – see section
A.10 End Tidal CO2 (Model 623xx).
Poor placement of nasal cannula in patient's nose.
;
not displayed.
2
78
BP

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents