Welch Allyn Atlas Operator's Manual page 97

Hide thumbs Also See for Atlas:
Table of Contents

Advertisement

PROBLEM:
Temperature display is inaccurate
RECOMMENDED ACTIONS:
Skin probe:
Verify proper placement on patient.
Eliminate drafts or other airflow near probe – cover with tape.
Oral, rectal, skin probe:
Replace probe.
Have unit repaired.
E.2 Atlas Monitor Error Messages
MESSAGE:
Altitude too high - CO
Altitude too low - CO
MEANING:
The ambient barometric pressure is too high or low for Atlas Monitor ETCO
RECOMMENDED ACTIONS:
If you are not above 1700 meters, or below 170 meters below sea level, this indicates a failure or
calibration error in the Atlas Monitor barometric sensing system. The unit must be repaired.
MESSAGE:
Battery depleted - shutdown imminent
MEANING:
The battery (models 622xx and 623xx) is drained. The system will shut down within one minute.
RECOMMENDED ACTIONS:
If you have AC power available, plug in the Atlas Monitor. The battery will recharge automatically,
whether the Atlas Monitor is turned on or off. The battery will return to 80% charge within 5 hours, and
full charge within 24 hours. You will see the
When operating on battery power, you can extend the battery life by reducing the number of Blood
Pressure measurements and printouts.
MESSAGE:
Battery is missing
MEANING:
The Atlas Monitor cannot detect the battery (models 622xx and 623xx). It has been removed, or there has
been a failure of the battery or its connections.
RECOMMENDED ACTIONS:
The battery must be replaced.
You can still operate the Atlas Monitor on AC power if available.
unavailable
2
unavailable
2
symbol lit above the ECG cable connection.
AC~
85
functions.
2