Calculating The Total Magnification And Field Of View Diameter - Leica M125 Manual

Hide thumbs Also See for M125:
Table of Contents

Advertisement

Calculating the Total Magnification and Field of View Diameter

Parameter
MO
Magnification of objective
ME
Magnification of eyepiece
z
Magnification changer position
q
Tube factor, e.g. 1.5× for coaxial incident light, 1.6× for 45°
ErgoTube™
r
Factor 1.25x if the planachromatic and planapochromatic
objectives of the MZ125/MZ16 are used on the MS5, MZ6,
MZ75 or MZ95
NFOV Field number of the eyepiece. Field numbers are printed on the
eyepieces: 10× = 21, 16× = 14, 25× = 9.5, 40× = 6.
Calculation example: magnification in binocular tube
MTOT VIS = MO × ME × z × q × r
or
1 × 25 × 4 × 1.5 × 1.25 = 187.5×
Leica M series
Example
MO
1× objective
ME
25×/9.5 eyepiece
z
Zoom position 4
q
Coaxial reflected light 1.5×, tube factor
r
Factor 1.25×
Calculation example: field of view diameter in the specimen
N FOV
∅ OF:
M O × z ×q ×r
Manual
112

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M205 cM165 cM205 a