Calculating Total Magnification And Field Diameter - Leica M-Series User Manual

Stereomicroscopes
Table of Contents

Advertisement

Notes
Total magnification and field diameter
M O
Magnification of objective
M E
Magnification of eyepiece
z
Magnification changer position
q
Tube factor, e.g. 1.5 for coaxial incident light,
45° Ergotube™ 1.6
r
Factor 1.25 if the planachromatic and planapochromatic
objectives of the MZ12
MS5, MZ6, MZ7
or MZ9
5
N FOV Field number of eyepiece.
The field numbers are printed on the eyepieces:
10 = 21, 16 = 14, 25 = 9.5, 40 = 6.
Example
M O
Eyepiece 1.0
M E
25 /9.5 eyepiece
z
Zoom position 4.0
q
Tube factor 1.5 (because of coaxial illuminator).
r
Factor 1.25
Magnification down binocular tube:
M TOT VIS = M O
Diameter of field of view:
40
/MZAPO are used on the
5
5
M E
z
q
N FOV
OF :
M O
z
q
r
r
1.0
25
9.5
1.0
4.0
1.5
4.0
1.5
1.25 = 187.5
= 1.3 mm
1.25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ms5Mz6Mz75Mz95Mz125Mzapo

Table of Contents