Keithley 706 Instruction Manual page 71

Table of Contents

Advertisement

1. Sets programmed first channel to 1.
2. Sets programmed last channel to 100.
3. Sets the interval time to 2OOmsec.
4. Sets the unit to the display mode.
5. Use the continuous scan mode.
6. Trigger the scan sequence on GET.
7. Provide a pause and clear statement.
Enter the following statements into the HP-85 to program the
Model 706 to scan channels 1 through 100 at a 200msec in-
terval rate.
PROGRAM
COMMENTS
10 REMOTE 718 (END LINE)
Model 706 set to remote.
20 CLEAR 7 (END LINE)
Clears Model 706.
3-30
30 OUTPU'I"718;
"DOFl LlOOW.2P2Tw".
(END LINE1
Programs the Model 706
for the above parameters.
40 TRIGGER 718
Provides GET trigger.
50 PAUSE
Program pause.
60 CLEAR 7
Clears Model 706.
70
END
End of program.
To start the program, press the RUN key on the HP-85.
After the RUN key is pressed the Model 706 scans the pro-
grammed channels at the programmed interval rate. To stop
the program simply press the CONT key. The program will
stop at the second channel when the CONT key is pressed.
To resume the program press the CONT key a second time
or press the RUN key. Upon pressing the RUN key or the
CONT key (second time) the Model 706 continues the pro-
gram.