4. Number of poles is 2.
5. Use single scan mode.
Use the steps outlined in Table 2-7 to program the Model 706
for the preceding parameters.
Upon the actuation of the START/STOP
button in step 10,
channel 5 is closed for 0.5 seconds (programmed interval
rate) and then opens. The Model 706 advances to the next
channel and closes it for the programmed interval rate and
then opens it.
The sequence continues up to and including the last pro-
grammed channel. At this point the Model 706 reverts to the
programmed first channel and turns off the START/STOP
LED. The single scan sequence is now complete.
Pressing the START/STOP
button during a single scan se-
quence stops the scan at the presently displayed channel.
The channel remains closed and turns off the START/STOP
LED. To continue the scanning sequence from the presently
displayed channel press the START/STOP
button.
To stop the scanning sequence and reset the Model 706 to
the programmed first channel (which is 5 in this example)
simply press the RESET button.
Example
3 Continuous
Scan-In
the continuous
scan
mode the scanning sequence described in example 2 repeats
continuously
until at least one of the following conditions are
met:
1. Power is interrupted.
2. During
the sequence,
the START/STOP
button
is
pressed.
3. RESET button is pressed.
4. PRGM button is pressed.
5. DATE button is pressed.
6. TIME button is pressed.
7. CHANNEL button is pressed.
In the continuous scan mode all the programmed channels
are scanned sequentially and the scan sequence is repeated
continuously.
The continuous scan is started by pressing the
START/STOP
button and it operates as follows:
1. The START/STOP
LED is turned on.
2. The programmed first channel is closed for the programm-
ed interval rate and then opened.
3. The Model 706 then advances to the next channel, closes
it for the programmed interval rate and then opens it.
4. This sequence is repeated up to and including the pro-
grammed last channel. At this point the Model 706 reverts
to the programmed first channel and repeats the sequence
continuously.
To help illustrate a continuous scan sequence we will choose
the following
parameters. For the sake of simplicity we will
use the Model 7056 general purpose scanner card.
1. First channel is 1.
2. Last channel is 100.
3. Interval time is 0.3 seconds.
4. Number of poles is 2.
5. Use continuous scan mode.
Use the steps outlined in Table 2-8 to program the Model 706
for the preceding parameters.
-
Table 2-7. Single Scan Example
Ster, 1 Action
I Comments
1.
Install the three Model 7056 scan-
Refer to paragraph 2.2.2.
ner cards into slots 1, 2
* and 3 of the Model 706.
2
Select appropriate line voltage
and turn on the Model 706.
3
Press PRGM, 6.
Selects PROGRAM 6.
4
Press 2, ENTER.
Selects the 2-pole mode. Refer to paragraph 2.9 for
more information concerning front panel programs.
z
Press 0, 0, 5, FIRST.
Programs channel 5 as the first channel.
Press 0, 2, 5, LAST.
Programs channel 25 as the last channel.
7
Press INTERVAL, 0, 0, 0, 5,
Programs the interval time for 0.5 seconds.
0, 0, ENTER.
8
Press CHANNEL, RESET.
Turns on the CHANNEL LED, selects the channel
mode for display and resets to the programmed first
channel.
9
Press SINGLE.
Selects the single scan mode and turns on the
SINGLE LED.
10
Press START/STOP.
Initiates the single scan sequence.