Keithley 706 Instruction Manual page 54

Table of Contents

Advertisement

Table 3-8. Hierarchy
of Command
Execution
I
Command
E**
S **
v**
Q **
H"
W*
Y"
B*
I***
C ***
N ***
z***
FYY"
L **+
A ***
RW,y+
Definition
Date Fromat
Time
Date
Alarm Time
Settle Time
Interval Time
Terminator
Channel
Channel Save
Close Channel
Open Channel
Channel Recall
Set First Channel
Set Last Channel
Set Number of Poles
Reset
*Independent
Commands
**Timer Commands
***Channel
Commands
3.5.1 Display Mode
The following five commands set the display mode for the
channel, time, date, interval or display message mode.
1. DO- The DO command sets the Model 706 for the channel
display mode.
2. Dl -The
Dl command sets the Model 706 for the interval
time display mode. The interval time is the programmed
time the channel is closed. The interval time can be altered
by programming inputs, refer to paragraph 3.5.8.
3. D2-The
D2 command sets the Model 706 for the time
display mode. The time in hours, minutes and seconds is
displayed, refer to paragraph 2.6 for time display format.
4. D3-The
03 command sets the Model 706 for the date
display mode. The date in month and day is displayed,
refer to paragraph 2.6 for date display format.
5. D4- The D4 command sets the Model 706 for the display
message mode. The display message mode allows the user
to display a customized message on the Model 706 scan-
ner's front panel display.
Programming
Example-To
display and transmit the chan-
nel data with a prefix and display the data on the HP%% CRT
enter the following program into the HP-85.
Notes
EO = American Date Format
El = international
Date Format
Set Time
Set Date
Set Alarm Time
Set Settle Time
Set Interval Time
IEEE Terminator Character
Display Channel
11,12,13,14,15...175 save a relay set up, IO clears all
buffers.
Close Channel
Open Channel
Zl, 22, 23, 24, Z5...275 recall a saved relay set up.
Overrides close.
Set First Channel
Set Last Channel
if changed First = 1 Last = last channel (100 to
500) and displays channel 1.
Opens all channels, displays first channel and
overrides close, open and recall.
PROGRAM
COMMENTS
10 REMOTE 718 (END
Sets the Model 706 for the
LINE)
remote mode.
20 CLEAR 7 (END LINE)
Clear the Model 706.
30 OUTPUT 718; "GODOX"
Display channel mode on
706 display and send chan-
nel date with prefix.
40 ENTER 718;AS (END
Obtain data string.
LINE)
50 DISP A$ (END LINE)
Display data string.
60 END (END LINE)
End of program.
Alter typing in the program, press the RUN key on the HP-85
to run the program. After the RUN key is pressed the channel
data with a prefix is displayed on the CRT as follows:
coo01 ,so
This program can be used for any of the prefix commands
(Gl.
As another
example we will
illustrate
the D4 (display
message) mode. Read the following
paragraphs and do the
programming example.
3-13

Advertisement

Table of Contents
loading

Table of Contents