9 Parts counting

With parts counting you can either count parts into a container or remove parts from
a container. To count a greater number of parts the average weight per part has to
be determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness. High reference must be selected for
small parts or parts with considerably different sizes.

9.1 Determination of the reference weight by weighing

Operation
Set your balance at zero
Tare the weighing container:
Determine the reference unit:
Place a known number of parts on the
balance as reference weight
If the "Weight" display is stable, enter
the number of parts via number keys.
Confirm within 5 sec:
The reference weight has been deter-
mined
The reference quantity is displayed in the
"Quantity" display
46
Display
z0
NET
a
z0
a
NET
5z0
a
NET
Piece weight
10
PRESET
Piece weight
5.00001
PRESET
10
CPB-BA-defsi-0611
Weight
Weight
Weight
Pcs