Electrode Stability Test; Meter Factory Reset - Thermo Scientific Orion Star A213 Reference Manual

Benchtop and portable rdo/do meters
Hide thumbs Also See for Orion Star A213:
Table of Contents

Advertisement

Electrode Stability Test

2. Press
3. Press
press f2 (yes).
8. The meter will display Stability Test Passed or Stability Test Failed. Press f1 (esc) or
measurement mode.

Meter Factory Reset

After a factory reset is performed, all of the meter settings will be reset to the factory defaults and the data log,
calibration log and methods will be deleted.
2. Press
3. Press
Thermo Scientific Orion Star A213 Benchtop and Star A223 and Star A323 Portable RDO/DO Meter Reference Guide
,
,
or
to highlight Diagnostics and press f3 (select).
or
to highlight Stability Test and press f2 (accept).
or
to highlight DO/RDO – Channel and press f2 (next).
dissolved oxygen and polarographic dissolved oxygen probes.
,
,
or
to highlight Diagnostics and press f3 (select).
or
to highlight Factory Reset and press f2 (accept).
f2 (yes) to reset the meter.
.
.
f2 (yes).
to return to the
33