Calibration and Adjustment
Weighing
Statistics
Calibrate balance
CAL-Extern
CAL-Intern
Report
Date and Time:
14-Jun-2012
Calibration function:
External calibration
Start calibration:
Start: manually
Calibration weight:
Set 2000.00 g
Calibration deviation:
Dev 0.02 g
Adjustment function:
External adjustment
Adjustment deviation:
Dev 0.00 g
72
User Manual for Secura, Quintix, Practum
Calibration/Adjustment using an
External Calibration Weight
An external calibration weight is required for this function.
Please note the tolerance of the calibration weight being used.
In verified balances with accuracy class K, external calibration/adjustment is
locked when the balances are used for legal metrology.
®
Secura
and Quintix
adjustment can only be accessed when the access switch is open.
t
Make sure that the weighing pan is empty.
t
A
Select
t
Select the Menu key at the bottom left of the display in the current application.
y The menu appears.
Mixing
Components
t
Select the
y The
Calibrate balance
t
Select
CAL-Extern
y Practum
t
Select the value for your calibration weight.
y A message appears on the display, prompting the user to place the calibration
weight on the weighing pan.
t
Place the calibration weight on the weighing pan.
y Calibration begins automatically.
y A report appears on the display (Secura
The report indicates the deviations identified during calibration. The report also
14:13
gives the result of the adjustment process.
t
<
Select
y The balance is now calibrated and adjusted.
: In verified balances with accuracy class k, external
®
to zero the balance.
button to access the calibration function.
CAL
window appears.
.
®
models: Calibration weight values appear in the display.
to close the report window.
®
and Quintix
®
only).