Keithley 2701 Service Manual page 142

Ethernet-based dmm / data acquisition system
Hide thumbs Also See for 2701:
Table of Contents

Advertisement

Model 2701 Service Manual
Table B-1 (Continued)
Remote calibration command summary
Command
:CALibration
:PROTected
:DC
:STEP9 <NRf>
:STEP10 <NRf>
:STEP11 <NRf>
:STEP12 <NRf>
:AC
:STEP1
:STEP2
:STEP3
:STEP4
:STEP5
:STEP6
:STEP7
:STEP8
:STEP9
:STEP10
:STEP11
:STEP12
:STEP13
:STEP14
:STEP15
NOTE: Upper-case letters indicated short form of each command. For example, instead of sending
":CALibration:PROTected:INITiate," you can send ":CAL:PROT:INIT."
1
DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in manufacturing calibration
mode.
Description
1MΩ 4-wire step.
10mA DC step.
100mA DC step.
1A DC step.
AC cal steps.
10mV AC at 1kHz step.
100mV AC at 1kHz step.
100mV AC at 50kHz step.
1V AC at 1kHz step.
1V AC at 50kHz step.
10V AC at 1kHz step.
10V AC at 50kHz step.
100V AC at 1kHz step.
100V AC at 50kHz step.
700V AC at 1kHz step.
100mA AC at 1kHz step.
1A AC at 1kHz step.
2A AC at 1kHz step.
1
1V AC at 3Hz step
.
1
1V AC at 1kHz step
.
Calibration Reference
B-3