SECTION 12—BATTERIES
Connecting Battery Cables
The use of rubber gloves and chemical goggles or face shields is recommended
when working with batteries.
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery
terminal(s)/post(s) at the same time. An electrical short may occur and serious per-
sonal injury or damage may occur.
Perform one (1) of the following methods for connecting the battery cable(s):
• Dual U1 Batteries ‐ Use direct mount method. Refer to FIGURE 12.5 and FIGURE 12.6.
• Dual Group 22NF Batteries with Mounting Holes in the Terminal(s)/Post(s) ‐ Use
direct mount method. Refer to FIGURE 12.5 and FIGURE 12.6.
• Dual Group 22NF Batteries without Mounting Holes in the Terminal(s)/Post(s) ‐ Use
battery clamp method. Refer to FIGURE 12.7, FIGURE 12.8 and FIGURE 12.9.
Direct Mount Method
NOTE: For this procedure, refer to FIGURE 12.5 and FIGURE 12.6.
1. Install battery terminal cap(s) onto battery cable(s):
A. Dual U1 Batteries ‐
i. ORANGE battery terminal cap onto RED battery cable.
ii. GREY battery terminal cap onto BLACK battery cable.
B. Dual Group 22NF Batteries ‐
i. RED battery terminal cap onto RED battery cable.
ii. BLACK battery terminal cap onto BLACK battery cable.
Battery Terminal Cap
NOTE: Only one battery cable and terminal cap shown for clarity. Both caps install in the same
manner.
Nutron® Series
WARNING
Battery Cable
FIGURE 12.5 Installing Battery Terminal Cap
55