Troubleshooting - Velleman Personal Scope HPS10 User Manual

Personal scope
Hide thumbs Also See for Personal Scope HPS10:
Table of Contents

Advertisement

_________________________________________________________________________________________________________________________________________________________
The screen remains blank or there is no signal:
No power supply
Batteries are discharged
Contrast adjustment is incorrect
Press RESET for at least 10 seconds
REMARK: Temporarily remove both batteries and adapter in case RESET is not helping.
RMS readout is incorrect:
Make sure that at least 1 and preferably even 2 periods are displayed.
The batteries are discharged.
No signal on the oscilloscope display:
Time/div setting is in the wrong position. Try 1ms or choose auto-setup mode.
The unit is in the Hold position
Trigger function is set in the "once" position
The programmed trigger level is not reached (choose "run" mode)
Y position is wrong or need calibration, see page 11
The input signal is too high, change the volt/div. setting or choose auto-setup.
Incorrect frequency readout:
An incorrect time/div. setting has been chosen. (start at 250ns/div)
Voltage readout does not correspond with the actual value:
The measuring probe is in the X10 position
The batteries are discharged
The zero reference is not set correctly for DC measurements.
RS232 Cable (Only for HPS40) :
3,5mm STEREO PLUG
7
2
3
24

TROUBLESHOOTING

*The pictures may differ depending on the type of PersonalScope
9P SUB-D FEMALE
2
3
7