For all AC measurements: Make sure that at least one or two periods are displayed or select the
auto-setup function.
•
You can choose "none" to hide readouts.
•
Depending on the selected screen layout one to four different meter readouts can be displayed.
•
At 1s/div timebase or slower, the readouts are forced to the instant information. 'Vs' (Fig. 25)
•
If no key is pressed during 10 seconds, the unit returns to its previous operating mode and the
cursorkeys are set for changing the timebase and sensitivity.
PROBE SETUP
G
Notes:
•
Automatically calculate the correct readouts depending the x1 or x10 probe setting.
•
An 'x10' symbol is displayed if this mode has been selected.
•
X10 measuring probes should be calibrated!
•
IMPORTANT: Set the measuring probe in the x10 position for measuring high voltages (>100Vp+dc)
Setting up a X10 measuring probe
When used in the X10 position, a measuring probe should always be
calibrated to the measuring instrument being used, in this case the
Personal Scope. (Fig. 26)
•
Open the battery cover.
•
Set the scope to X10 position (probe x1/x10 key)
•
Set the voltage per division to 1V
•
Set the time per division to 0.1ms
•
Select AC for the input.
Use the probe to perform measurements at the pre-selected point under the battery cover. Adjust the
trimmer of the measuring probe in order to obtain a square wave signal with a top that is as flat as
possible. (Fig. 27)
The pictures may differ depending on the type of PersonalScope
Press the 'Probe x1/x10'-key to set the measurements
Fig 25
accordingly the x1/x10-probe setting. (Fig. 25)