TROUBLESHOOTING

The screen remains blank or there is no signal :

No power supply
Batteries are discharged
Contrast adjustment is incorrect
Press RESET for at least 10 seconds
REMARK: Temporarily remove both batteries and adapter in case
RESET is not yet operational.

RMS readout is incorrect :

Make sure that at least 1 and preferably even 2 periods are displayed.
The batteries are discharged.

No signal on the oscilloscope display :

Time/div setting is in the wrong position. Try 1ms or choose auto-
setup mode.
The unit is in the Hold position
Trigger function is set in the "once" position
The programmed trigger level is not reached (choose "run" mode)
Y position is wrong
The input signal is too high, change the volt/div. setting or choose
auto-setup.

Incorrect frequency readout:

An incorrect time/div. setting has been chosen. (start at 2µs)
Voltage readout does not correspond with the actual value :
The measuring probe is in the X10 position
The batteries are discharged
The zero refrence is not set correctly for DC measurements.
Take into account that the RMS readout only measures the AC
voltage