Caster Plates; U.anti-Tip Tubes - Rear And Front - Sunrise Medical Zippie GS Owner's Manual

Hide thumbs Also See for Zippie GS:
Table of Contents

Advertisement

Z I P P I E G S
T. CASTER PLATES
1. Angle Adjustment
a) Loosen the eccentric bolts (B) that secure the caster assembly.
b) Place a large right triangle against the flat surface and the front surface of the caster
bearing housing. This will align the caster stem rotational axis perpendicular to the flat
surface.
c) Align and retighten the two eccentric bolts.
d) Repeat the process for the other caster.
2. Angle Adjustment: Caster Plates with Transit Option
a) Loosen the bolts (C) that secure the caster assembly.
b) Place a large right triangle against the flat surface and the front surface of the caster
bearing housing. This will align the caster stem rotational axis perpendicular to the flat
surface.
c) Align the eccentric washers and retighten the bolts.
d) Repeat the process for the other caster.

U.ANTI-TIP TUBES - REAR AND FRONT

NOTE: Sunrise recommends anti-tip tubes for all wheelchairs.
1. To Insert Anti-Tip Tubes into Receivers
a) Press the anti-tip release button (D) on the anti-tip tube so that the release pin (E) is
drawn inside.
b) Insert the tube into the anti-tip receiver.
c) Turn the anti-tip tube until release pin is positively engaged in the receiver mounting
hole (F).
d) Insert second anti-tip tube the same way.
2. To Adjust Anti-Tip Height
a) Press the anti-tip wheel release button (G) so the release pin is drawn inside.
b) Raise or lower to one of the three pre-drilled holes.
c) Turn the anti-tip tube until release pin is positively engaged in the receiving hole (H).
d) Adjust the second anti-tip tube wheel the same way. Both wheels should be exactly the
same height.
930377 Rev. F
25
26
27
28
29
27
28
29
18
25
B
26
27
C
28
F
E
D
29
H
G