Sets Number of Scanning Cycles
number
ARM:COUN<
through the channel list per
Default is one scanning cycle.
Selects the Trigger Source
TRIG:SOUR
defines the trigger source to
advance the scan. Default is
TRIG:SOUR?
queries the current trigger source.
Sources are:
BUS = *TRG
EXT =
Event In connector
Hold triggering
HOLD =
IMM =
Automatic advance
Selects Continuous Scanning Cycles
Use
INIT:CONT ON
continuous scanning cycles. Default made is
fixed number of scans, with number set by
number
ARM:COUN<
Selects Scan Mode
SCAN:MODE
sets the RF multiplexers for
NONE
,
VOLT
,
RES
and
have no effect on RF multiplexer
RES
operation.
FRES
sets "paired-channel"
scanning.
Selects the Scan List
channel_list
SCAN<
be scanned using the trigger source set by
the
command.
TRIG:SOUR
Enables Scanning
starts scanning for any trigger source
INIT
and closes the first channel in the channel list
Advances the Scan
TRIG
can be used to advance the scan for
or
TRIG:SOUR BUS
command is not functional when
or
TRIG:SOUR EXT
32
Understanding the HP E1366A/E1367A RF Multiplexers
sets 1 to 32767 scans
>
INIT
command.
TRIG:SOUR IMM.
via HP-IB interface
or
INIT:CONT 1
to enable
>.
, or
FRES
.
NONE
,
VOLT
defines channels to
>
The
TRIG:SOUR HOLD.
TRIG:SOUR IMM
is set.
Figure 4-1. Scanning Channels Commands
,
.