Immunity Test
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the USBCam2 is used exceeds the
applicable RF compliance above, the USBCam2 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the USBCam2.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
Table 7. Recommended Separation Distance Between Portable and Mobile RF
PLEASE NOTE: The USBCam2 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or user of the USBCam2 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
USBCam2 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of the transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
24
IEC 60601 Test
Compliance
Level
Communications Equipment and the USBCam2
Separation distance according to the frequency of the
150 kHz to 800 MHz
d=1.2 x √P
0.12
0.38
1.2
3.8
12.0
B1051061 Rev. –
Level
transmitter (m)
800 MHz to 2.5 GHz
d= 2.3 x √P
0.23
0.73
2.30
7.3
23.00
Guidance
USBCam2 User Guide