4.1.4. Pausing and Resuming Live Video
NOTE: The following steps apply to the use of USBCam2 with CDR DICOM software.
To set up your handpiece Freeze Frame button for Freeze/Unfreeze mode, do the
following:
Click the Setup button on the Video Capture window
Select option for Freeze/Unfreeze
Click OK to save settings
To use your handpiece in this mode, depress the Freeze Frame button once (Figure 1, item
2) to pause live video. Depressing the button again will resume streaming video.
To capture an image when live video is paused, click the Take button on the Video Capture
window.

Figure 5. Video Window Setup and Capture Buttons

4.2. Using Camera Sheaths

Disposable sheaths are provided with each USBCam2 System as an effective measure for
infection control. Use a new sheath for each intra-oral use, and then dispose it of properly.
When the sheath is inserted over the handpiece, notice that the upper part of the sheath
(covering the lens) has both a clear and frosted side. Make sure the CLEAR side of the
sheath is facing the camera lens.