C-2. Troubleshooting Table

Item
Description
1
Camera LEDs illuminate and remain on
immediately after connecting USB cable,
whether software is running or not.
2
Camera turns on and off intermittently.
3
Cannot select 640 x 480 output size in Video
Format dialog box.
4
Spots are visible on video image.
5
Extraoral images appear grainy or blurry.
6
Camera options are missing from the
DirectShow Video dialog box.
7
Corrupted video image is displayed when using
the camera with an Ultra Port USB 2.0 PCI card,
ALI chipset.
8
Momentary blip occurs recurrently during video
exam.
9
Image appears nearly completely red during
video exam.
USBCam2 User Guide
This is normal camera operation. To turn off LEDs, press
the Power button on the camera, or place the camera in its
handpiece.
Check USB cable connection at camera and at PC
connectors.
Camera is not connected to High-Speed USB 2.0 port.
Reconnect camera to appropriate port.
Check camera lens for spots and follow the instructions in
the Handpiece and Lens cleaning section of this
document (Section 5.2). Also check the "fit" of the
camera sheath over the lens. Camera sheath should be on
tight enough so there are no wrinkles across the lens
section.
Camera is strictly intended for intraoral use.
Camera driver was not installed correctly. Disconnect
camera from USB cable, uninstall USBCam2 driver, and
follow the instructions in the Software Setup section of
this document (Section 3).
Replace the USB card with one that is compatible with
the camera.
Change to a different USB port.
Toggle between the Fixed Balance / Auto White Balance
options in the Camera properties dialog box.