Protective Measures; Sheaths; Handpiece And Lens; Computer - Schick USBCam2 User Manual

Hand-held intraoral camera
Hide thumbs Also See for USBCam2:
Table of Contents

Advertisement

5. Protective Measures

5.1. Sheaths

The USBCam2 system uses disposable sheaths (Schick P/N 6310120) to ensure proper
infection control. Sheaths are intended for single-use only. Do not use the camera for
intraoral use until a new sheath has been properly fitted to the handpiece. After each intra-
oral use, remove the sheath and dispose of it properly. Then, clean the outer surfaces of the
camera using mild cleaning solution (mild soap and water) and sterile wipes. A new sheath
must be used for each patient.
Disposable sheaths are provided with every USBCam2 system. In the U.S., additional
sheaths can be ordered from Patterson Dental Supply, Inc. by calling 1-800-328-5536.
International customers should contact the authorized international dealer for Schick
Technologies products in their country or region.

5.2. Handpiece and Lens

IMPORTANT! Be sure to disconnect the camera from its USB cable and port before
performing any cleaning procedures. Before cleaning the lens, be certain that it is free of
any abrasive contaminants. This will help to avoid the formation of scratches while cleaning.
To disinfect the handpiece, use a 70% Ethanol solution, administered by saturating spray
or moist towelettes. Do not soak the handpiece and be sure to dry it completely.
To clean the handpiece, use a clean lint-free cloth and any of the following solutions. Do
not soak the handpiece and be sure to dry it completely.
Mild soap and water
Isopropyl alcohol (70%)
Isopropanol (>15%) / quaternary ammonium compounds combination
(<1%)
To clean the lens surface, moisten a soft cotton swab dipped in rubbing (isopropyl)
alcohol. Gently wipe the lens surface end-to-end in straight lines without applying
pressure; do not scrub. Allow the alcohol to evaporate from the lens surface. Repeat these
steps, as needed, until the lens surface is clean.

5.3. Computer

To avoid cross contamination, follow the cleaning instructions provided by your computer
manufacturer and implement them as part of your normal routine for ensuring proper
sterilization and disinfectant of tools in your dental practice.