Performing a Control Test

You need the meter, a test strip, and control solution Level 1 or Level 2. The control level is printed on the bottle label.
1
Check the meter display
for missing segments.
All segments should look
like the picture.
5
Remove the bottle cap.
Wipe the tip of the bottle
with a tissue.
Squeeze the bottle until a
tiny drop forms at the tip.
2
or
Select the control solution
to test. You will enter the
level later in the test.
Touch the drop to the
front edge of the yellow
window of the test strip
until you see flash.
Do not put control
solution on top of the
test strip.
3
Insert the test strip into
the meter in the direction
of the arrows.
Place the meter on a flat
surface.
6
When you see flash,
there is enough control
solution in the test strip.
Control Tests
4
Make sure the code
number on the display
matches the code number
on the test strip container.
If you miss seeing the
code number, remove the
test strip and reinsert it
into the meter.
The control result, the
bottle symbol, and a
flashing L appear on the
display.
Do not remove the test
strip yet.
13