Counting Mode (Pcs) - A&D HR-250AZ Instruction Manual

Hr-az hr-a series analytical balance
Table of Contents

Advertisement

6-3 Counting Mode (PCS)

This is the mode to determine the number of objects in a sample based on the standard sample unit
mass. The unit mass means the mass of one piece of the sample. The smaller the variables in each
sample unit mass are, the more accurate the counting will be. The balance is equipped with the
Automatic Counting Accuracy Improvement (ACAI) function to improve the counting accuracy.
Notes
For counting, use samples with a unit mass of at least ten or more times the minimum
weighing value of the balance.
If the sample unit mass variable is too large, it may cause a counting error.
To improve the counting performance, use the ACAI function frequently or divide the samples
into several groups and count each group.
Selecting the counting mode
1 Press the MODE key to select
Storing a sample unit mass
2 Press the SAMPLE key to enter the sample unit mass
storing mode.
Even in the storing mode, pressing the MODE key will
switch to the next mode.
3 To select the number of samples, press the SAMPLE key
several times. It may be set to 10, 25, 50 or 100.
Note
A greater number of samples will yield more accurate
counting result.
4 Place a container on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare). The
number specified in step 3 appears.
e.g.: 25 0 is displayed if 25 is selected in step 3.
5 Place the number of samples specified on the pan. In this
example, 25 pieces.
6 Wait for the stabilization indicator to turn on. Press the
PRINT key to calculate and store the unit mass. The
balance displays
count samples with this unit mass.
To improve the accuracy of the unit mass, go to step 8.
(counting mode).
(counting mode) and is set to
16
Weighing
pan
Container
Sample
(25 pieces)
Place sample
to be counted
The counting result

Hide quick links: