A&D HR-250AZ Instruction Manual page 42

Hr-az hr-a series analytical balance
Table of Contents

Advertisement

Class
Item and Parameter
btpr
Data bit, parity bit
Crlf
Terminator
5if
type
Serial
Data format
interface
t-Up
Timeout
erCd
AK, Error code
ldin
d5 fnc
Liquid density input
Density
d5
function
Selection of a
mltMLT
Programmable-unit (Multi-unit)
Unit
Unit
id
ID number
apf
Application function
Aap fnc
5taf
Application
Statistical function
mode output items
C5 in 1 *
Internal mass value correction: Method 1
C5 in 2 *
Internal mass value correction: Method 2
Caution
The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate.
When a refresh rate of 20 times/second is used, set the baud rate to 4800 bps or higher.
0
7 bits, even
1
7 bits, odd
2
8 bits, none
0
CR LF
1
CR
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
5
CSV format
0
No limit
1
1 second
0
No output
1
Output
0
Water temperature
1
Liquid density
0
Density measurement of a solid
1
Density measurement of a liquid
Sets an arbitrary coefficient.
Refer to "5. WEIGHING UNITS".
Refer to "11. ID NUMBER AND GLP REPORT"
0
Normal weighing mode
1
Capacity indicator
2
Statistical calculation mode
0
Number of data, sum
Number of data, sum, maximum,
1
minimum, average, range
(maximum-minimum)
Number of data, sum, maximum,
minimum, average, range
2
(maximum-minimum), standard
deviation, coefficient of variation
Number of data, sum, maximum,
minimum, average, range
3
(maximum-minimum), standard
deviation, coefficient of variation,
relative error
Refer to "8. CALIBRATION".
Factory setting
Only for the HR-AZ series
*
:
40
Description
CR: ASCII code 0Dh
LF: ASCII code 0Ah
Refer to "10-6 Description of the
Item "Data Format"".
Selects the wait time to receive a
command.
AK:ASCII code 06h
Available only when the density
mode is selected and stored in the
unit selection mode.
Refer to "13 Density Measurement".
Available only when programmable-
unit mode is selected.
Displayed only when the
function switch (internal mass
value correction switch) is set
to 1.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents