BC Biomedical SA-2010S User Manual

Electrical safety analyzer with patient simulator
Hide thumbs Also See for SA-2010S:

Advertisement

ELECTRICAL
SAFETY ANALYZER
WITH PATIENT SIMULATOR
SA-2010S
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SA-2010S and is the answer not in the manual?

Questions and answers

Summary of Contents for BC Biomedical SA-2010S

  • Page 1 ELECTRICAL SAFETY ANALYZER WITH PATIENT SIMULATOR SA-2010S USER MANUAL...
  • Page 3: Table Of Contents

    BC BIOMEDICAL SA-2010S TABLE OF CONTENTS WARNINGS ......................... 3 DESCRIPTION......................4 LAYOUTS ........................6 TESTING........................10 TEST MATRIX ....................17 ECG WAVEFORMS ..................... 18 MANUAL REVISIONS....................19 WARRANTY......................... 19 SPECIFICATIONS ....................... 20 APPENDIX A -- LABEL ....................22 NOTES......................... 23...
  • Page 4 POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES. CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL, INC. PO BOX 25125 9415 GENTRY AVE ST. LOUIS, MO 63125...
  • Page 5: Warnings

    Do not leave the device under test (DUT) connected to this unit for extended periods. It is intended for short duration testing within the current limits and duty cycles specified. Manual SA-2010S Copyright © 2006 www.bcgroupintl.com Made in the USA...
  • Page 6 ELECTRICAL SAFETY ANALYZER WITH PATIENT SIMULATOR The Model SA-2010S is a Microprocessor based Electrical Safety Analyzer with a built in Patient Simulator. It allows for a multitude of tests to be performed on a device using the same unit and lead connections. It provides a full Electrical Safety Analyzer, as well as, ECG Simulation with four waveforms with constant QRS duration and six machine performance testing waveforms.
  • Page 7 PATIENT SIMULATOR: • ECG: 30, 60, 120 AND 240 BPM • SINE: 10, 60 AND 100 HZ • SQUARE: 0.125 AND 2.0 HZ • TRIANGLE: 2 HZ • HIGH LEVEL OUTPUT (1 V p-p) • AMPLITUDE ACCURACY: +/- 2% • FREQUENCY ACCURACY: +/- 0.5%...
  • Page 8: Layouts

    LAYOUT This section looks at the layout of a SA-2010S and gives descriptions of the elements that are present. The unit is controlled by 9 light touch keys. They allow the user to perform all the functions available in the system. There is an audio click when a key is depressed. Note that some keys are locked out based on the current function setting.
  • Page 9: Description

    Display The main information in the system is presented in the 3½ digit LCD display. This data is provided as simple meter readings with the units indicated by one of the four LEDs on the right of the display. The units LEDs will automatically change to the necessary range based on the function selected.
  • Page 10 Analyzer Test Receptacle Control There are four keys and 8 LEDs in the Analyzer Test Receptacle Control Section. They allow the manual control of the power connections that are made to the DUT. Internally, a series of relays are switched by the microprocessor based on the keys that are depressed. The LEDs indicate the current state of the power connections to the Receptacle.
  • Page 11 High Level Output (+) A High Level snap is located on the side of the unit for connecting to the high level ECG output signal (1 Volt p-p ). The connection is between High Level (+) and RL (-). Connectors There are two connectors for test cables on the unit.
  • Page 12: Testing

    SAFETY ANALYZER TESTING WARNING All connections to patients must be removed before connecting the Device Under Test (DUT) to the Analyzer. A serious hazard may occur if the patient is connected when testing with the Analyzer. Do not connect any leads from the patient directly to the Analyzer or DUT.
  • Page 13 Mains Voltage With the Mains Voltage function selected, the display will show the Voltage that is present on the incoming power lines. This is measured from Line to Neutral. Insure that this value is within the specifications for the equipment that is to be tested. Also note that the voltage may drop when the DUT is turned on.
  • Page 14 Earth/Ground Leakage Current With the Earth/Ground function selected and the Ground-Open, the display will show the leakage current in the ground wire of the DUT. The display is in µAmps and will read from 0 to 1999. Selecting this function automatically opens the connection to Earth/Ground and passes any leakage current through a 1000 Ohms load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load key.
  • Page 15 Lead to Earth/Ground Leakage With the Lead to Earth/Ground function selected, the display will show the leakage current between the selected Patient Lead and Earth/Ground. The display is in µAmps and will read from 0 to 1999. Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead or all of the leads.
  • Page 16 Lead Isolation WARNING The Analyzer applies 120 or 240 VAC plus 10% to the Patient leads or External test leads during the Isolation test. Although this is current limited by a 121K Ohms internal resistor, per standard test specifications, care should be taken to prevent contact with this voltage.
  • Page 17 Point to Point Measurements The unit has the ability to measure Leakage Current, Isolation Leakage Current and Resistance between two points, utilizing two test leads. These tests are separated because they use a slightly different setup than the previous tests. Point to Point Leakage Current With the External function selected, the display will show the leakage current between the test leads.
  • Page 18 Point to Point Resistance With the Earth Resistance function selected, the display will show the resistance between the two test leads. The display is in hundredths of Ohms and will read to 19.99 Ohms. Remove any device plugged into the Analyzer Test Receptacle. The test requires two Black (Kelvin) test cables--one in the Chassis Socket and the other in the External Socket.
  • Page 19: Test Matrix

    Test Matrix GROUND POLARITY NEUTRAL POWER Closed Forward Closed Closed Closed Forward Closed Open Closed Forward Open Closed Closed Forward Open Open Closed Reverse Closed Closed Closed Reverse Closed Open Closed Reverse Open Closed Closed Reverse Open Open Open Forward Closed Closed Open...
  • Page 20: Ecg Waveforms

    Performance Waves Sine The SA-2010S sends out Sine waves of 10, 60 and 100 Hz. With ECG Waveforms function selected, use the arrow keys under the waveform LEDs to select the desired Sine wave. The display is unused in this mode and will show over range.
  • Page 21: Manual Revisions

    MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 DT7338 Preliminary Manual Rev 02 DT7338CC Photos Updated LIMITED WARRANTY WARRANTY : BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED.
  • Page 22: Specifications

    SPECIFICATIONS SCALES & RANGES 85 – 265 VAC, 50/60 Hz, RMS LINE VOLTAGE +/- 3% of reading, +/- 1 digit 0 – 19.99 Amps, RMS DUT CURRENT +/- 5% of reading, +/- 1 digit EARTH/GROUND 0-19.99 Ohms RESISTANCE +/- 1% of reading, +/- 1 digit LEAKAGE CURRENT 0-1999 µAmps, RMS DC and 25 to 1KHz...
  • Page 23 OUTPUT LEAD 1 1.75 mV LEAD 2 2.75 mV LEAD 3 1.00 mV LEAD TO LEAD 1000 Ohms IMPEDENCE LEAD TEST < 1000 Ohms IMPEDENCE ACCURACY AMPLITUDE +/- 2 % Lead II FREQUENCY 0.5 % PHYSICAL 8.625 x 5.5 x 1.625 Inches ENCLOSURE (219 x 140 x 41 mm) ABS Plastic...
  • Page 24: Appendix A -- Label

    APPENDIX A -- LABEL The following is a representation of the label found on the back of the SA-2010S:...
  • Page 25: Notes

    NOTES...