en | Installation
NOTICE!
The backup battery will now charge up. The specified standby
time is available when the battery is fully charged, after 16
hours. If the Carephone 62 is stored and disconnected, connect
it to the main power supply at least once every 6 months, to
allow the battery to charge.
5.3.2

IP connection

Refer to the user manual of the IP module.
5.3.3

VoIP connection

When using a VoIP connection, be aware of the following
limitations: network availability, priority of the emergency call,
power failures.
5.3.4

Connection configuration

NOTICE!
In the event of an emergency, the unit must be able to send out
an emergency call even if a phone call is in progress at the time.
The Carephone 62 must interrupt the phone call.
To check whether an emergency call can be sent:
1.
2.
If the setup is correct, the phone call will be interrupted.
CAUTION!
With a private branch exchange (PABX) there is no guarantee
that a phone call will be interrupted by the emergency call.
5.4

Deinstallation

To switch off the Carephone 62 and the backup battery, remove
the power plug from the unit.
NOTICE!
If you remove the power plug from your power outlet, the unit
announces "power failure". The unit will switch to battery
operation.
F.01U.141.162 | V3 | 2012.11
Make a phone call as you normally would.
When the phone call is in progress, press the red button.