3.9.6.
Multiple Load Connection with Distribution Terminals
If remotely located output distribution terminals are used, the power supply output
terminals should be connected to the distribution terminals by pair of twisted and/or
shielded wires. Each load should be separately connected to the remote distribution
terminals (see Fig.3-6).
If remote sensing is required, the sensing wires should be connected to the distribution
terminals or at the most critical load.
Power
Supply
-Rem Sense
-Local Sense
+Local Sense
+Rem Sense
Fig. 3-6: Multiple loads connection with distribution terminal
3.9.7.
Grounding Outputs
Either the positive or negative output terminals can be grounded. To avoid noise probems
caused by common-mode current flowing from the load to ground, it is recommended to
ground the output terminal as close as possible to the power supply chassis ground.
Always use two wires to connect the load to the power supply regardless of how the
system is grounded.
+V
-V
There is a potential shock hazard at the sense connector when using
a power supply with a rated output voltage greater than 40V. Local
sense and remote sense wires should have a minimum insulation
rating equivalent or greater than the maximum output voltage of the
power supply. Ensure that the connections at the load end are
shielded to prevent accidental contact with hazardous voltages.
83-530-000 Rev J
Distribution Terminal
+V
-V
WARNING
38
+
Load #1
-
+
Load #2
-
+
Load #3
-