Figure 8. Measuring Polarization Index and Dielectric
Absorption Ratios
16
or
bbw10f.eps
Using the Compare Function (Model
1507)
Use the Compare function to set a pass/fail compare level
for the insulation measurements. To use the Compare
function:
Press the C button to select the desired compare
1.
value. You can choose from 100 kΩ, 200 kΩ, 500 kΩ,
1 MΩ, 2 MΩ, 5 MΩ, 10 MΩ, 20 MΩ, 50 MΩ, 100 MΩ,
200 MΩ, and 500 MΩ.
2.
Perform insulation tests as described earlier in this
manual.
3.
The green pass indicator will appear if the measured
value is greater than the selected value.
Press and hold the C button for 1 second to
4.
disable the Compare function. The pass indicator will
turn off when you start a new test or choose a new
compare value.