2150EX Area Velocity Flow Module
Section 2 Preparation and Installation
Figure 2-19 Connecting the AV Sensor
Important Information
Regarding "X" Marking
Figure 2-20 AV2150EX Sensor Cable Labels
2-26
3. Align and insert the connector. The sensor release will click
when the sensor connector is fully seated.
4. Connect the two caps together.
The ATEX labeling on the sensor's serial tag shows a number
ending in "X". The X marking indicates that there are special
conditions that must be met to ensure safety, as explained on
page vi in the front of this manual.
The AV2150EX sensor is labeled and X-marked for special
usage conditions in order to prevent static electricity. Avoid
conditions that may generate a static charge, such as rubbing
the AV2150EX with static producing cloth.
X-Marking
Sensor Release
CAUTION
WARNING
ELECTROSTATIC HAZARD
DO NOT RUB
CLEAN WITH DAMP CLOTH ONLY
Caps