Model UT60D: OPERATING MANUAL
Measurement Operation (2)
l
When AC voltage measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test.
DC Voltage Measurement
The DC Voltage ranges are: 340.0mV, 3.400V, 34.00V,
340.0V and 1000V. To measure DC voltage, connect
the Meter as follows:
1.
Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
2.
Set the rotary switch to V
is default or press BLUE button to select DC
measurement mode.
3.
Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l
In each range, the Meter has an input impedance
of 10M
measurement errors in high impedance circuits. If
the circuit impedance is less than or equal to 10k ,
the error is negligible (0.1% or less).
l
When DC voltage measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test.
(see figure 4)
red
Hz; DC measurement
.
This loading effect can cause
18
black
(figure 4)