Co Sensor Calibration; High Sensor Calibration; Material Required; Procedure - Bacharach PCA3 Operation And Maintenance Manual

Portable combustion analyzer siegert version
Hide thumbs Also See for PCA3:
Table of Contents

Advertisement

PCA
3 Manual
6.11. CO
Sensor Calibration
HIGH
This procedure spans the optional carbon monoxide high sensor (4,001 to
20,000 ppm) to a known gas level.

6.11.1. Material Required

Calibration Kit
Gas Cylinder

6.11.2. Procedure

1.
If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
Use the ▲▼ buttons to highlight CO-HI, and then press ENT to display
2.
the CALIBRATE CO-HI screen.
3.
Attach a 4,000 ppm CO cylinder to the regulator of the calibration fixture
shown in Figure 6-12.
4.
Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the concentration on the CO cylinder.
80
(P/N 0024-7059)
(4,000 ppm CO in Air, P/N 0024-1155)
Figure 6-12. Gas Sensor Calibration Equipment
NOTE:
"Measured" is the current CO reading, while
"Applied" is a known CO level that will be applied for
calibration purposes.
0024-9473 Rev 3