5.2 Ambient Temperature Calibration

Material Required:
Procedure:
1. Enter the Calibrate Mode as described in Section 5.1. Observe that
"AMBIENT" is displayed at the top of the display; if not, repeatedly
press ENTER until it appears.
2. Use the ▲ and ▼ buttons to adjust the displayed value to match
the reading of a calibrated thermometer.
3. Press ENTER to store the displayed value and move to the next
calibration screen, or hold down ENTER for 2 seconds to store the
displayed value and display the CO screen, or press the I/O button
to exit the Calibration Mode and turn OFF the analyzer without
saving the changes.
5.3 CO Sensor Zero & Calibration
Material Required:
If you are primarily performing flue-gas testing, calibrate the Monoxor III
with 500 ppm CO calibration gas. If the analyzer will be primarily used for
ambient testing, then you may wish to consider 100 ppm CO calibration gas.
Procedure:
1. With the analyzer sampling fresh air, enter the Calibration Mode as
described in Section 5.1. Then repeatedly press the ENTER button
until "CO" appears in the upper left side of the display.
20
Monoxor III
Calibrated Thermometer
• Cylinder of 100 ppm (P/N 0051-1994) or
500 ppm (P/N 0024-0492) CO calibration gas
• Calibration Kit, (P/N 0024-7059)
Instruction 0019-9326