Bacharach Fyrite Pro Operation & Maintenance Manual

Bacharach Fyrite Pro Operation & Maintenance Manual

Combustion gas analyzer
Hide thumbs Also See for Fyrite Pro:
Table of Contents

Advertisement

Fyrite
Pro
®
Combustion Gas Analyzer
Instruction 24-9388
Operation & Maintenance
Rev. 6 – July 2009
Product Leadership • Training • Service • Reliability

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Fyrite Pro and is the answer not in the manual?

Questions and answers

Summary of Contents for Bacharach Fyrite Pro

  • Page 1 Fyrite ® Combustion Gas Analyzer Instruction 24-9388 Operation & Maintenance Rev. 6 – July 2009 Product Leadership • Training • Service • Reliability...
  • Page 2 Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of this Product or parts thereof that are defective in title.
  • Page 3: Table Of Contents

    Fyrite Pro Contents Ordering Information ................iii 1.0 INTRODUCTION ..................1 1.1 The Fyrte Pro..................1 1.2 Operatonal Overvew ..............2 2.0 TECHNICAL CHARACTERISTICS ..........3 3.0 SETTING UP THE ANALYZER ............5 3.1 Prelmnary Steps ................5 3.2 Checkng & Replacng the Batteres ..........5 3.3 Connectng the Probe and Thermocouple ........6 3.4 Front Panel Push Buttons ..............8 3.5 Setup Mode ..................9 3.5.1 Enterng Setup ..............9 3.5.2 Selectng Temperature Unts ..........9 3.5.3 CO Channel Setup (Model 125E only) ......
  • Page 4 Fyrite Pro 4.9 Clearng Saved Test Data .............24 4.10 Turnng OFF the Analyzer & CO Purge ........24 4.11 Resettng the Mcroprocessor ............25 4.12 Run Mode Screens ................26 4.13 Screen Descrptons ..............28 4.13.1 Warm-Up Screens Includng O Sensor Output ..28 4.13.2 Fuel Screen ..............28 4.13.3 TA (Stack Temperature) & Draft Screen ....29 4.13.4 Draft Zero Screen ............29 4.13.5 Stack & EFF Screen .............30 4.13.6 Rato of CO to CO Screen (Model 125E only) .... 30 4.13.7 qA & Excess Ar Screen ..........31 4.13.8 TL & Eta Screen ............31 4.13.9 CO & CO Ar Free Screen (Model 125E only) .....
  • Page 5: Ordering Information

    Fyrite Pro Ordering Information Model 122E 125E Part Number 24-7276 24-7220 Measurements Oxygen   Stack Temperature   Ambient Temperature   Second Thermocouple Channel   (Combustion Air Temperature) Carbon Monoxide (CO)  Draft   Calculations qA (Stack Loss) ...
  • Page 6 Fyrite Pro Notes: Instruction 24-9388...
  • Page 7: Introduction

    Fyrite Pro 1.0 INTRODUCTION 1.1 The Fyrite Pro The Fyrte Pro seres of nstruments are hand-held combuston effcency analyzers that are desgned for on-demand samplng of resdental furnac- es and applances. Ths manual contans nformaton on all models of the Fyrte Pro. Please dsregard any nformaton that does not pertan to your model. WARNINGS! This analyzer is not intended to be used on a continuous basis. This analyzer does not have an audible alarm, and is not intended to be used as a safety device.
  • Page 8: Operational Overview

    Fyrite Pro 1.2 Operational Overview Pressng the I/O button turns the analyzer ON. Note that there s a warm- up perod of 60 seconds before the unt can be used for testng. After completon of the warm-up perod, choose a fuel code that corre- sponds to the fuel beng burned by the applance beng tested, and then press the eNTeR button to place the analyzer nto ts Run Mode. Begn testng by nsertng the analyzer’s probe tube nto the flue-gas stream of the applance under test. The analyzer wll montor the flue gas and dsplay measured and calculated values that are relatve to the combuston process. These values are dsplayed on the analyzer’s LCD, and are chosen for dsplay by pressng the Increment () and Decrement () buttons. Pressng the hOlD button durng a test wll freeze all measured and calculat- ed values at ther current levels. Pressng the eNTeR button resumes testng. Pressng the hOlD button twice durng a test wll dsplay the Prnt/ Memory Menu Screen. Usng the  and eNTeR buttons, the operator can choose to prnt the current readngs, save the current readngs n memory, recall and vew saved readngs from memory, clear all the memory, or prnt all saved readngs. A backlght enables an operator to read the dsplay n dmly-lt areas. Pressng the lIGhT button turns the backlght ON and OFF. The backlght wll automatcally turn OFF after 10 mnutes of keyboard nactvty. A power saver functon wll shut the analyzer OFF after 20 mnutes of keyboard nactvty. The power saver functon s dsabled f the detected O value s less than 17.9%, or for analyzers wth a CO sensor the CO value s more than 50 ppm. Press the I/O button to turn the analyzer OFF. Note that there s a 5 sec- ond delay before the analyzer actually turns OFF, durng whch tme the unt can be kept ON by pressng the eNTeR button. For analyzers that contan a CO sensor, there s a CO purge feature that keeps the analyzer’s...
  • Page 9: Technical Characteristics

    Fyrite Pro 2.0 TeChNICAl ChARACTeRISTICS The Analyzer Directly Measures and Displays: • Combuston Ar (TL) ........0 to 100 °C (32 to 212 °F) • Flue Gas O content ........0.0 to 20.9% O • Flue Gas Temperature (TA) ......–18 to 537 °C (0 to 999 °F) • Flue Gas CO content .....0 to 2,000 ppm CO (Model 125E only) • Draft ..............–9.99 to 67 mb/HPa (–9.99 to 27 nWC) The Analyzer Computes and Displays: (when the measured oxygen level s below 17.9%) • qA (Stack Loss) ..........0 to 99% • ETa ..............0 to 109.6% • Combuston Effcency ........0.1 to 99.9% • Flue Gas CO Ar-Free content...
  • Page 10 Fyrite Pro Power Requirements ......Four dsposable ‘AA’ Alkalne bat- teres (an optonal AC adapter s avalable that elmnates the need for batteres) Operating Time (using batteries) . .. Up to 18 hours contnuous (pump runnng and backlght off) Warm Up Time ........60 seconds Memory ..........Up to 10 sets of readngs can be saved n memory. When memory s full, the next readng saved wll overwrte the oldest readng. Display ..........4 Dgt, 2 Lne, 7-segment Lqud Crystal Dsplay Front Panel Controls ....... Sx push button swtches (Refer to Secton 3.4) Accuracy: • Oxygen..........±0.3% O wth a typcal flue gas concentraton of CO • Carbon Monoxde ......
  • Page 11: Setting Up The Analyzer

    Fyrite Pro 3.0 SeTTING UP The ANAlYZeR 3.1 Preliminary Steps Before usng the analyzer . . . • Check batteres (Secton 3.2) • Connect probe to analyzer (Secton 3.3) • Check setup (Secton 3.5) 3.2 Checking & Replacing the Batteries Install fresh batteres as descrbed below. Check the analyzer for suffcent charge pror to each use. Replace the batteres f the low-battery symbol appears n the lower rght corner of the screen. To replace the bat- teres: 1. Remove battery cover from back of analyzer. 2. If old batteres are nstalled, remove them and properly dscard. 3. Observng the polarty markngs nsde the battery compartment, nstall four ‘AA’ Alkalne batteres as shown n Fgure 3-1. 4. Replace battery cover. Remove battery cover by pushing down on cover and sliding it outward Figure 3-1.
  • Page 12: Connecting The Probe And Thermocouple

    Fyrite Pro 3.3 Connecting the Probe and Thermocouple Attach the probe and hose assembly to the analyzer as follows (Fgure 3-2): 1. Push the yellow banded quck-connect Flue Gas Hose (gvng a slght twst) onto the GAS nlet fttng. 2. Push the blue banded quck-connect Draft Hose (gvng a slght twst) on the “+” pressure fttng. 3. Push the Flue Gas Thermocouple nto the T-STACK jack Important: DO NOT force the thermocouple connector into its jack. The connection tabs are different sizes, allowing the connector to fit in only one way.
  • Page 13 Fyrite Pro Figure 3-2. Connecting the Probe and Hose Assembly Instruction 24-9388...
  • Page 14: Front Panel Push Buttons

    Fyrite Pro 3.4 Front Panel Push Buttons Note that a push button may perform several functons, dependng on the analyzer’s model number and what screen s beng dsplayed at the tme. I/O • Toggles the analyzer ON and OFF. • Places the analyzer nto ether ts Setup or Calbraton Mode when used n conjuncton wth the eNTeR or hOlD button. • Scrolls up through the dsplay screens durng a test.  • Causes the dsplayed value to ncrease or change whle n the Calbraton, Setup, or Prnt/Memory Menu Screen. • Scrolls down through the dsplay screens durng a test.  • Causes the dsplayed value to decrease or change whle n the Calbraton, Setup, or Prnt/Memory Menu Screen. e NTeR • Enters the Run Mode (starts a combuston test) from the Fuel Screen. • Unfreezes the dsplay after pressng the hOlD button. • Dsplays the Fuel Screen when held down for 2 seconds whle n the Run Mode. • Performs the acton selected n the Prnt/Memory Menu Screen. • Stores the dsplayed value and automatcally steps to the next screen when pressed durng calbraton or setup. • Places the unt nto ts Run Mode when held down for 2 seconds whle n the Calbraton Mode. • Aborts turn-off and keeps the analyzer turned ON when pressed durng the 5 second turn-off-delay perod. • Aborts the CO purge functon at turn-off when the measured CO level s above 50 ppm.
  • Page 15: Setup Mode

    Fyrite Pro 3.5 Setup Mode The analyzer s preset at the factory for the parameters shown below, but can be changed as descrbed n ther assocated sectons. Function Parameter To Change Fuel Natural Gas (F1) Secton 4.13.2 Temperature Unt °C Secton 3.5.2 CO Channel Auto Zero (A2) Secton 3.5.3 Draft (Pressure) Secton 3.5.4 Clock Not Intalzed Secton 3.5.5 thru 3.5.7 Prnter Protocol IrDA Secton 3.5.8 3.5.1 Entering Setup 1. Wth the analyzer turned OFF, press and hold down the hOlD button. 2. Press the I/O button. 3. Release both buttons. 4. The analyzer s now n ts Setup Mode. Refer to Sectons 3.5.2 thru 3.5.8 for nformaton on how to set up each parameter. 3.5.2 Selecting Temperature Units The Temperature Setup Screen s labeled “Unt.”...
  • Page 16: Co Channel Setup (Model 125E Only)

    Fyrite Pro 3.5.3 CO Channel Setup (Model 125E only) The CO Channel Setup Screen s labeled “CO.” 1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly press the eNTeR button untl “CO” s dsplayed. 2. Press ether the  or  button untl the desred parameter s d splayed. - no Sensor dsabled. Use ths functon to turn off the CO chan- nel f the sensor s bad or not nstalled to avod the CO s ensor error message from appearng. - 2Ero Manual Zero. When the CO channel s set to manual zero, the analyzer does not zero the CO sensor to ambent cond- tons durng start up. - A2 Auto Zero. Wth the CO channel set to auto zero, the CO sensor s zeroed to the ambent CO level durng start up. Important: When using this mode the analyzer must be turned ON in fresh air; otherwise, incorrect CO readings will occur.
  • Page 17: TMe Setup

    Fyrite Pro 3.5.5 Time Setup There are two Tme Setup Screens, one for hours and the other for mn- utes. Two bars appear above the segments beng changed. Tme s ds- played n a 24 hour format. 1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly press eNTeR untl the frst Tme Setup Screen s dsplayed—the one wth two bars over the hour dgts. 2. Press ether the  or  button untl the correct hour value s d splayed. 3. Press eNTeR to move the selecton bars over the mnute dgts. 4. Press ether the  or  button untl the correct mnute value s dsplayed. 5. Press eNTeR to move to the next Setup Screen, or I/O to ext setup. 3.5.6 Date Setup There are two Date Setup Screens, each labeled “DAtE”. The frst screen sets the month (second two dgts) whle the second screen sets the day (frst two dgts). 1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly press eNTeR untl the frst “DAtE” Screen s dsplayed. 2. Press ether the  or  button untl the correct month s dsplayed. 3. Press eNTeR to change the ‘Day’ value. 4. Press ether the  or  button untl the correct day s dsplayed.
  • Page 18: Year Setup

    Fyrite Pro 3.5.7 Year Setup The Year Setup Screen s labeled “yEAr”. 1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly press eNTeR untl “yEAr” s dsplayed. 2. Press ether the  or  button untl the correct year s dsplayed. 3. Press eNTeR to move to the next Setup Screen, or I/O to ext setup. 3.5.8 Selecting Printer Protocol The analyzer can be set up for ether an HP or IrDA type prnter. 1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly press eNTeR untl one of the followng screens s dsplayed. 2. Press ether the  or  button to select the desred prnter protocol. 3. Press eNTeR to move to the next Setup Screen, or I/O to ext setup. 3.5.9 Exiting the Setup Screen Press the I/O button at any tme to ext the Setup Mode and turn OFF the analyzer. Note that the last dsplayed parameter s automatcally saved n memory. Instruction 24-9388...
  • Page 19: Operation

    Fyrite Pro 4.0 OPeRATION 4.1 Performing CO and Draft Testing 4.1.1 Analyzer Turn On and Warm Up 1. Connect the probe and make sure that the analyzer s properly set up per Secton 3.0. 2. Turn ON the analyzer by pressng s I/O button and observe that the followng Warm-Up Screens are dsplayed. These screens show the analyzer’s model number (Pro E 122 or Pro E 125), the unt’s software revson, and fnally the remanng warm-up tme counted down from 60 seconds, ncludng a measure- ment of the O sensor’s output level. Tip: An O error will occur when the oxygen sensor’s output drops to between 80 and 90. Consider replacing the oxygen sensor when its output level drops below 100.
  • Page 20: MeasurNg Co (Model 125E Only)

    Fyrite Pro 4.1.2 Measuring CO (Model 125E only) 1. Turn ON the analyzer and allow t to warm-up per Secton 4.1.1. Important: If the CO channel is set up for auto zero, then the analyzer must be turned ON in fresh air; otherwise, incorrect CO readings will occur. Refer to Section 3.5.3. When the CO channel is set up for manual zero, and if the CO reading is higher than zero when sampling fresh air, then manually zero the CO channel per Section 5.6, Steps 1 thru 4.
  • Page 21: Performing A Combustion Test

    Fyrite Pro 4.2 Performing a Combustion Test 4.2.1 Analyzer Turn On and Warm Up Important: The probe must be at room temperature before performing the following steps. 1. Connect the probe, and f used, the combuston-ar thermocouple to the bottom of the analyzer per Secton 3.3. 2. Before turnng ON the analyzer, poston the probe or the optonal combuston-ar thermocouple to measure the temperature of the burner’s combuston ar by performng one of the followng: • When NOT USING the Combustion-Air Thermocouple – Before turnng ON the analyzer, ts probe must be located n the area contanng the burner’s combuston-ar supply. If the burner s usng room ar, then smply place the probe wthn the room. In the case of a hgh-effcency furnace where combuston ar s...
  • Page 22: SamplNg PoNt

    Fyrite Pro 4.2.3 Sampling Point Forced Air Furnace – When testng atmospherc burner or gravty vented, forced ar heatng equpment wth a clamshell or sectonal heat exchanger desgn, test each of the exhaust ports at the top of the heat ex- changer. The probe should be nserted back nto each of the exhaust ports to obtan a flue gas sample, before any dluton ar s mxed n. Hot Water Tank – Domestc hot water tanks wth the ‘bell’ shaped draft dverter on top can be accurately tested by nsertng the probe tp drectly nto the top of the fre tube below the dverter. 80% Efficiency Fan Assist or Power Vented – Combuston testng of fan assst or power vented, furnaces/bolers should be done through a hole drlled n the vent mmedately above the nducer fan. 90% Efficiency Condensing – Condensng furnaces/bolers can be tested through a hole drlled n the plastc vent ppe (when allowed by the manufac- turer or local authorty of jursdcton) or taken from the exhaust termnaton. Atmospheric or Gravity Vented Boiler – Bolers, whch have a ‘bell’ shaped draft dverter drectly on top, should be tested drectly below the dverter through a hole drlled n the vent connector. Atmospheric Burner or...
  • Page 23 Fyrite Pro 80% eff. Fan Assist or Power Vented Furnace/Boiler 90% eff. Condensing Furnace/Boiler Atmospheric or Gravity Vented Boiler Instruction 24-9388...
  • Page 24: PerformNg A CombustOn EffCEncy Test

    Fyrite Pro 4.2.4 Performing a Combustion Efficiency Test 1. After turnng ON the analyzer and selectng the approprate fuel, the TA & Draft Screen (refer to Secton 4.13.3) should be dsplayed. 2. Poston the probe n the flue-gas stream to obtan the hottest TA (stack temperature) readng. Locating the highest stack temperature is very important for accurate calculations of qA, Eta, and efficiency. 3. Burner-servce procedures can now begn. Use the  and  buttons to scroll through the analyzer’s other dsplay screens (refer to Sec- ton 4.12). The analyzer readngs wll update contnuously showng changes n burner performance. Note: When a calculation cannot be made because of im- proper data (i.e., oxygen level above 17.9%), four dashes “- -...
  • Page 25: OperatNg TPs

    Fyrite Pro 4.3 Operating Tips • When an analyzer s brought n from a cold vehcle, let t warm up slowly to mnmze condensaton. Temperatures below freezng wll not damage the analyzer; however, brngng a cold analyzer nto a warm, humd envronment may cause condensate to form nsde the case. • If the CO channel (Model 125E only) s set up for Auto Zero (refer to Secton 3.5.3), ensure that the analyzer s samplng fresh ar when turned ON. Pullng a flue-gas sample through the analyzer durng ts warm-up perod wll not damage the analyzer, but t wll result n n- correct CO readngs. Also note that a CO sensor error wll occur f the detected CO level s above 50 ppm durng warm-up. • When measurng flue-gas, note that flue-gas condensate s acdc and very corrosve. It s mportant not to allow the analyzer’s nternal com- ponents to become soaked n condensate for long perods of tme. • Before each use, nspect the flter element of the water-trap / flter as- sembly. Replace the flter f t looks drty. Refer to Secton 5.8. • When samplng flue-gas, keep the analyzer above the water-trap, and keep the trap n a vertcal poston. Ths wll maxmze the effectve- ness of the trap and keep lqud condensate from beng drawn drectly nto the analyzer. • When lqud condensate s seen nsde the water trap, empty the trap before t becomes full. Refer to Secton 5.8. • It s recommended that the analyzer be purged after takng a flue-gas measurement before turnng t OFF. Once the probe s removed from the stack, dsconnect the hose assembly from the bottom of the analyz- er and let the pump run for 10 mnutes or so to completely remove any remanng flue gases and dry any condensate from nsde the sensor chamber. • When storng the analyzer, t’s a good dea to empty the water trap and leave t open to further dry t out. • Calbrate the analyzer every 6 months to ensure ts accuracy. Instruction 24-9388...
  • Page 26: UsNg The BacklGht

    Fyrite Pro 4.4 Using the Backlight The LCD can be read n dmly-lt areas by pressng the lIGhT button. The backlght automatcally turns OFF after 10 mnutes of keyboard nactv- ty, but can be turned OFF at any tme by agan pressng the lIGhT button. 4.5 Using the Probe A rgd stanless steel probe wth handle, connected to a flexble hose wth ntegral water-trap / flter can be used to draw a gas sample nto the ana- lyzer from the room, grlles, dffusers, and furnace flues. The hose and probe assembly can be detached from the analyzer when the operator desres to sample wthout the probe (.e., when samplng for CO). 4.6 Saving Test Data in Memory Up to 10 ndvdual sets of test data can be saved n memory as follows: Note: When memory is full, the next reading saved will over- write the oldest reading.
  • Page 27: OpenNg And VEwNg Saved Test Data

    Fyrite Pro 2. Press the  button once to dsplay the Save Screen. The number shown n ths screen represents the memory locaton (1 thru 10) to whch the current test data wll be saved. 3. Press eNTeR to save the test data and return to the Hold Mode, or press hOlD to return to the Hold Mode wthout savng. 4.7 Opening and Viewing Saved Test Data Perform the followng to open and vew saved test data: Note: If no test data has been saved, the option to open the memory for viewing will not be available.
  • Page 28: PrNtNg Test Data

    Fyrite Pro 4.8 Printing Test Data Turn ON the prnter. Refer to the printer’s instruction manual for detailed operation and maintenance information. If not already done, set the prnter parameters as follows: • Data: 8 bts • Baud: 9600 • Party: None • Handshakng: X-on/X-off Algn the prnter wth the top of the analyzer as shown n Fgure 4-1. The next step n the prntng procedure depends on whether the operator desres to prnt only the current test data, or all stored test data startng wth the most recent. Perform the approprate procedure below: Print Current Test Data 1. Wth the analyzer n ts Run Mode, press the hOlD button twice to dsplay the Prnt Screen. 2. Press eNTeR to start prntng. Print All Test Data 1. Wth the analyzer n ts Run Mode, press the hOlD button twice...
  • Page 29 Fyrite Pro � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �...
  • Page 30: Clearing Saved Test Data

    Fyrite Pro 4.9 Clearing Saved Test Data Clear all saved test data as follows: Note: If no test data has been saved, the option to clear memory will not be available. 1. Wth the analyzer n ts Run Mode, press the hOlD button twice to dsplay the Prnt Screen. 2. Press ether the  or  button to scroll to the Clear Screen, and then press eNTeR to dsplay the Clear All Screen. 3. Press eNTeR agan to clear memory and return to the Hold Mode, or press hOlD to return to the Hold Mode wthout clearng memory.
  • Page 31: Resetting The Microprocessor

    Fyrite Pro 4.11 Resetting the Microprocessor If the analyzer ‘locks up’ and cannot be turned OFF, reset the mcroproces- sor by removng one of the batteres for at least 5 seconds. Instruction 24-9388...
  • Page 32: Run Mode Screens

    Fyrite Pro 4.12 Run Mode Screens Fgures 4-2 and 4-3 show the order n whch the screens are dsplayed by pressng the eNTeR and  buttons after the analyzer warms-up and enters ts Run Mode. Refer to Secton 4.13 for a detaled descrpton of each screen. Fuel Select: F1 = Natural Gas F2 = Propane F3 = Light Oil F4 = City Gas � � � � � Stack Temperature & Draft Draft Zero Stack Temperature &...
  • Page 33 Fyrite Pro Fuel Select: F1 = Natural Gas F2 = Propane F3 = Light Oil F4 = City Gas � � � � � Stack Temperature & Draft Draft Zero Stack Temperature & efficiency Ratio of CO to CO qA &...
  • Page 34: Screen DescrPtOns

    Fyrite Pro 4.13 Screen Descriptions 4.13.1 Warm-Up Screens Including O Sensor Output As soon as the analyzer s turned ON, a seres of Warm-Up Screens are dsplayed. These screens show the analyzer’s model number (Pro E 122 or Pro E 125), the unt’s software revson, and fnally the remanng warm- up tme counted down from 60 seconds, ncludng a measurement of the O sensor’s output level. At the end of warm-up, the Fuel Screen s dsplayed. Note: If any errors were detected during warm-up, the Sensor Error Screen (Section 4.13.11) is displayed. Tip: An O error will occur when the oxygen sensor’s output drops to between 80 and 90.
  • Page 35: Ta (Stack Temperature) & Draft Screen

    Fyrite Pro 4.13.3 TA (Stack Temperature) & Draft Screen The TA readng s the stack temperature measured at the probe tp. The draft readng s the pressure measured at the probe tp as referenced to the analyzer’s “–” pressure port. Front Panel Button Functions: – Dsplays next screen (refer to Secton 4.12)  – Dsplays prevous screen (refer to Secton 4.12)  – Freezes dsplay (press eNTeR to unfreeze) hOlD eNTeR – Hold for 2 seconds to dsplay Fuel Screen lIGhT – Toggles backlght ON/OFF – Turns analyzer OFF (5 second delay) I/O 4.13.4 Draft Zero Screen The Draft Zero Screen allows the user to manually zero the draft channel when the pressure readng s anythng other than zero when measurng ambent room pressure. To zero the draft channel, remove all tubng from the analyzer’s “+” and “–” ports and press the eNTeR button. Front Panel Button Functions: – Dsplays next screen (refer to Secton 4.12)
  • Page 36: Stack & Eff Screen

    Fyrite Pro 4.13.5 Stack & EFF Screen The STACK readng s the temperature measured at the probe tp. The EFF (Effcency) readng s a calculaton of what percentage of energy present n the fuel s beng converted nto usable heat based on the fuel’s hgh heatng value. Front Panel Button Functions: – Dsplays next screen (refer to Secton 4.12)  – Dsplays prevous screen (refer to Secton 4.12)  – Freezes dsplay (press eNTeR to unfreeze) hOlD eNTeR – Hold for 2 seconds to dsplay Fuel Screen lIGhT – Toggles backlght ON/OFF – Turns analyzer OFF (5 second delay) I/O 4.13.6 Ratio of CO to CO Screen (Model 125E only) The COC2 Screen dsplays a readng that s the calculaton of ppm CO dvded by %CO as detected n the flue-gas stream. Note: The display does not show a decimal point in front of the reading.
  • Page 37: Qa & Excess AR Screen

    Fyrite Pro 4.13.7 qA & Excess Air Screen � The qA readng s a calculaton of stack loss n percent. The EXCESS AIR (LAMBDA) readng s the calculaton of the percentage of extra ar that s avalable n the combuston chamber above the theoret- cal amount needed for perfect combuston. Front Panel Button Functions: – Dsplays next screen (refer to Secton 4.12)  – Dsplays prevous screen (refer to Secton 4.12)  – Freezes dsplay (press eNTeR to unfreeze) hOlD eNTeR – Hold for 2 seconds to dsplay Fuel Screen lIGhT – Toggles backlght ON/OFF – Turns analyzer OFF (5 second delay) I/O 4.13.8 TL & Eta Screen The TL readng s the combuston-ar temperature that s used for ef- fcency and qA calculatons. Ths temperature s measured by ether the optonal T-AIR thermocouple on a contnuous bass, or by the T-STACK thermocouple durng start-up (refer to Secton 4.2.1). If neither thermo- couple s nstalled when the analyzer s turned ON, then ths temperature readng comes from a temperature sensor located nsde the analyzer.
  • Page 38: Co & Co AR Free Screen (Model 125E Only)

    Fyrite Pro 4.13.9 CO & CO Air Free Screen Model 125E only) The CO readng s the measurement of how much carbon monoxde n ppm s present n the flue-gas stream. The CO AIR FREE (undlute) readng s a calculaton of the exact ppm concentraton of carbon monoxde f oxygen were 0.0% n accordance wth ANSI standard Z21.1. CO x 20.9 Ar Free 20.9 – O 2 measured Front Panel Button Functions: – Dsplays next screen (refer to Secton 4.12)  – Dsplays prevous screen (refer to Secton 4.12)  – Freezes dsplay (press eNTeR to unfreeze) hOlD eNTeR – Hold for 2 seconds to dsplay Fuel Screen lIGhT – Toggles backlght ON/OFF – Turns analyzer OFF (5 second delay) I/O 4.13.10 CO & O...
  • Page 39: Sensor Error Screen

    Fyrite Pro 4.13.11 Sensor Error Screen An O sensor error s dsplayed f the analyzer determnes durng the warm-up cycle that the oxygen sensor’s output s too low for t to be usable. However, n the extreme condton when the O sensor has no output, a sen- sor error wll not occur. Instead, the O readng wll be 0.0 as dsplayed n the CO & O Screen. Tip: To avoid O errors from occurring, consider replac- ing the O sensor when its output level drop below 100 as displayed during warm-up (refer to Section 4.13.1).
  • Page 40 Fyrite Pro Notes: Instruction 24-9388...
  • Page 41: Calibration & Maintenance

    Fyrite Pro 5.0 CAlIBRATION & MAINTeNANCe Important: Fresh batteries should be installed, or use the optional AC adapter to power the analyzer during calibra- tion. The unit should be allowed to stabilize at room temper- ature for at least 2 hours before proceeding with calibration.
  • Page 42: The DSplay Segments

    Fyrite Pro 5.2 Ambient Temperature Calibration Material Required: Calbrated Thermometer Procedure: 1. Enter the Calbraton Mode as descrbed n Secton 5.1. Then re- peatedly press the eNTeR button untl “AMBIENT” appears at the top of the dsplay. 2. Use the  and  buttons to set the dsplayed value to match the readng of a calbrated thermometer at room temperature. 3. Press eNTeR to store the dsplayed value and move to the next calbraton screen, or hold down eNTeR for 2 seconds to store the dsplayed value and enter the Run Mode, or press the I/O button to ext the Calbraton Mode and turn OFF the analyzer wthout sav- ng the changes. 5.3 Combustion Air Thermocouple Calibration Material Required: Thermocouple Smulator Procedure: 1. Attach the thermocouple smulator to the analyzer’s Combuston Ar Thermocouple (T-AIR, see Fgure 3-2). 2. Enter the Calbraton Mode as descrbed n Secton 5.1. Then re- peatedly press the eNTeR button untl “TL” appears at the top of the dsplay. 3. Set the smulator to 0.0 °C (32.0 °F); then use the  and  buttons to set the dsplayed value to match the smulator’s value.
  • Page 43: Stack Thermocouple Calibration

    Fyrite Pro 5. Repeat Steps 3 and 4 as necessary untl the unt s readng correctly at both temperatures. 6. Press eNTeR to store these calbraton values and move to the next calbraton screen, or hold down eNTeR for 2 seconds to store these calbraton values and enter the Run Mode, or press the I/O button to ext the Calbraton Mode and turn OFF the analyzer wthout savng the changes. 7. Remove thermocouple smulator from analyzer. 5.4 Stack Thermocouple Calibration Material Required: Thermocouple Smulator Procedure: 1. Attach the thermocouple smulator to the analyzer’s Stack Thermo- couple connector (T-STACK, see Fgure 3-2). 2. Enter the Calbraton Mode as descrbed n Secton 5.1. Then re- peatedly press the eNTeR button untl “TA” appears at the top of the dsplay. 3. Set the smulator to 0 °C (32 °F); then use the  and  buttons to set the dsplayed value to match the smulator’s value. 4. Set the smulator to 300 °C (572 °F); then use the  and  buttons to adjust the dsplayed value to match smulator’s value. 5. Repeat Steps 3 and 4 as necessary untl the unt s readng correctly at both temperatures. 6. Press eNTeR to store these calbraton values and move to the next calbraton screen, or hold down eNTeR for 2 seconds to store these...
  • Page 44: Oxygen Sensor Zero

    Fyrite Pro 5.5 Oxygen Sensor Zero Material Required: • Cylnder of 100% Ntrogen, P/N 9550-0049 • Calbraton Kt, P/N 24-7059 Procedure: 1. Wth the analyzer samplng fresh ar, enter the Calbraton Mode as descrbed n Secton 5.1. Then repeatedly press the eNTeR button untl “O ” appears n the lower-left sde of the dsplay. 2. Allow pump to run and sample fresh ar for at least 1 mnute. 3. Use the  and  buttons to set the dsplay to 20.9%. 4. Set up the Calbraton Kt wth 100% N as descrbed n the nstruc- tons suppled wth the kt. 5. Connect the tubng of the Calbraton Kt to the GAS nlet of the analyzer; then adjust the regulator for approxmately 2 SCFH of excess flow (see Fgure 5-1). 6. After the analyzer has stablzed (2 to 3 mnutes,) use the  and  buttons to set the dsplayed value to 0.0%. 7. Dsconnect tubng from analyzer and turn off gas flow. 8. Allow the analyzer’s pump to run untl the O readng returns to 20.9. If necessary, use the  and  buttons to readjust the readng to 20.9. Repeat Steps 3 thru 7 to verfy the zero adjustment. 9. Press eNTeR to store the new calbraton values and move to the...
  • Page 45: Co Sensor Zero & CalBratOn (Model 125E Only)

    Fyrite Pro 5.6 CO Sensor Zero & Calibration (Model 125e only) Material Required: • Cylnder of 100 ppm (P/N 51-1994) or 500 ppm (P/N 24-0492) CO calbraton gas • Calbraton Kt, P/N 24-7059 To mprove the accuracy of the CO readng, we suggest that f the analyzer wll be prmarly used for flue gas testng, then calbrate usng 500 ppm CO. If the analyzer, however, wll be prmarly used for ambent testng, then calbrate usng 100 ppm CO. Procedure: 1. Wth the analyzer samplng fresh ar, enter the Calbraton Mode as descrbed n Secton 5.1. Then repeatedly press the eNTeR button untl “CO” appears n the upper-left sde of the dsplay. 2. Allow the pump to run and sample fresh ar for at least 1 mnute. 3. Use the  and  buttons to set the dsplayed value to 0 ppm. 4. Do one of the followng: a. End ths procedure and save the new zero value by holdng down the eNTeR button for 2 seconds; after whch, the analyzer enters ts Run Mode. b. Contnue wth Step 5 to span the CO sensor on a known concen- traton of carbon monoxde.
  • Page 46 Fyrite Pro 8. Press eNTeR to store the new calbraton values and move to the next calbraton screen, or hold down eNTeR for 2 seconds to store the new calbraton values and enter the Run Mode, or press the I/O button to ext the Calbraton Mode and turn OFF the analyzer wthout savng the changes. 9. Dsconnect tubng from analyzer and turn off gas flow. LEAVE TOP PORT OPEN Parts Shown: 1. Gas Cylinder 2. Regulator* 3. Tubing* 4. Tee* 5. Flowmeter* 6. Fitting, Gas* * Contained in Calibration Kit Figure 5-1. Calibration Kit Hookup...
  • Page 47: Draft Calibration

    Fyrite Pro 5.7 Draft Calibration Material Required: • Bellows • Mcromanometer Range: ±20 mb (±8 n. H O column) Accuracy: ±0.025 mB (±0.01 n. H O column) Procedure: 1. Set the draft unts to “mB” per Secton 3.5.4. 2. Enter the Calbrate Mode per Secton 5.1, and then press the eNTeR button untl “mB” appears on the rght-hand sde of the dsplay. 3. If necessary, zero the draft channel by usng the  and  buttons to adjust the readng to 0.00 ±0.02. 4. Set up the analyzer and ts test equpment per Fgure 5-2. 5. Use the bellows to apply a postve pressure of 10.00 mB (4.00 n. O), and then use the  and  buttons to adjust the analyzer’s dsplayed value to match the readng on the manometer. 6. Remove pressure and repeat Steps 3 thru 5 untl the desred read- ngs are obtaned. 7. Press eNTeR to store the new calbraton values and move to the next calbraton screen, or hold down eNTeR for 2 seconds to store the new calbraton values and enter the Run Mode, or press the I/O button to ext the Calbraton Mode and turn OFF the analyzer wthout savng the changes. Parts Shown: 1.
  • Page 48: Water Trap / FLter MaNtenance

    Fyrite Pro 5.8 Water Trap / Filter Maintenance The Water Trap / Flter Assembly removes water condensate from the gas sample, and also prevents soot from contamnatng the nternal compo- nents of the analyzer. Drain the water condensate after every test. Procedure: 1. Pull off the Inlet End Cap usng a slght twstng moton. 2. Pour out all water condensate; then reassemble trap. Replace the Filter element when dirty. Material Required: • Flter Element, P/N 07-1644 Procedure: 1. Pull off the Outlet End Cap usng a slght twstng moton. 2. Remove and dscard old flter. 3. Install new flter; then reassemble trap. FILTER ELEMENT...
  • Page 49: Sensor Replacement

    Fyrite Pro 5.9 O Sensor Replacement Be sure to perform all of the checks lsted n Secton 4.13.11 to ensure that the O sensor needs to be replaced. Material Required: • Oxygen Sensor, P/N 24-8106 • #1 Phllps Screwdrver Procedure: 1. Dsassemble the analyzer as follows: a. Remove the battery cover and the batteres, uncoverng one of the cover hold-down screws. b. Remove and set asde all four cover hold-down screws. c. Wth the analyzer on ts back, remove the front cover, layng t face down to the left of the body. d. Carefully remove the crcut board, slppng off the battery con- nector on top, and then layng the crcut board face down n the top cover. 2. Slp off the oxygen sensor’s electrcal connector from the crcut board. Then push down; twst counterclockwse; then pull the oxy- gen sensor out of ts socket (see Fgure 5-4). Tip: To obtain a better grip on the oxygen sensor, it may be necessary to remove the screw that secures the sensor socket to the case.
  • Page 50: Co Sensor Replacement (Model 125E Only)

    Fyrite Pro 5.10 CO Sensor Replacement (Model 125e only) Be sure to perform all of the checks lsted n Secton 4.13.11 to ensure that the CO sensor needs to be replaced. Material Required: • CO Sensor, P/N 24-7265 • CO Sensor Gasket, P/N 24-1112 • #1 Phllps Screwdrver Procedure: 1. Dsassemble the analyzer as follows: a. Remove the battery cover and the batteres, uncoverng one of the cover hold-down screws. b. Remove and set asde all four cover hold-down screws. c. Wth the analyzer on ts back, remove the front cover, layng t face down to the left of the body. d. Carefully remove the crcut board, slppng off the battery con- nector on top, and then layng the crcut board face down n the top cover. 2. Gently pull CO sensor out of ts socket (see Fgure 5-5). 3. Properly dspose of the old CO sensor (see the nstructon sheet that comes wth the new sensor). 4. It s recommended that the CO sensor gasket be replaced at the same tme as the sensor. Remove the old gasket and dscard. Re- move the paper backng from the new gasket and adhere t to the case n the same poston as the old one.
  • Page 51: Pump Assembly Replacement

    Fyrite Pro 5.11 Pump Assembly Replacement Material Required: • Pump, P/N 24-3048 • #1 Phllps Screwdrver Procedure: 1. Dsassemble the analyzer as follows: a. Remove the battery cover and the batteres, uncoverng one of the cover hold-down screws. b. Remove and set asde all four cover hold-down screws. c. Wth the analyzer on ts back, remove the front cover, layng t face down to the left of the body. d. Carefully remove the crcut board, slppng off the battery con- nector on top, and then layng the crcut board face down n the top cover. 2. Slp off the pump motor’s electrcal connector from the crcut board. 3. Unscrew the pump’s hold down clamp and remove t from the pump (see Fgure 5-4). Make note of how the pump wrng s routed. 4. Make note of how the tubng connects to the pump; then carefully remove tubng from pump. 5. Remove the old pump and dscard. 6. Install the new pump and renstall the tubng, takng care not to pnch or crmp the tubng. Also be sure pump wrng s routed as was noted n Step 3. 7. Reassemble the analyzer.
  • Page 52 Fyrite Pro CO SENSOR GASKET (Model 125E only) OXYGEN SENSOR & SOCKET PRESSURE SENSOR PUMP ON PCB HOLD DOWN CLAMP Figure 5-4. Inside Case Components Instruction 24-9388...
  • Page 53 Fyrite Pro BATTERY (RED WIRE) SENSOR (Model 125E only) OXYGEN SENSOR (RED WIRE) PRESSURE SENSOR PUMP (RED WIRE) T-AIR T-STACK Figure 5-5. PCB Components Instruction 24-9388...
  • Page 54 Fyrite Pro Notes: Instruction 24-9388...
  • Page 55: Parts & Service

    Fyrite Pro 6.0 PARTS & SeRVICe 6.1 Replacement Parts Description Part No. Carbon Monoxde Sensor ................24-7265 Carbon Monoxde Sensor Gasket ...............24-1112 Oxygen Sensor w/ Gasket ................24-8106 Oxygen Sensor Gasket ................24-1111 Probe Stop w/ Thumbscrew ...............19-3037 Pump Assembly ..................24-3048 Water Trap / Flter Assembly, Complete ............24-1107 Water Trap / Flter Assembly, Flter Element (pack of 3) ......07-1644 6.2 Accessories STANDARD: Battery, “AA” Alkalne ................204-0004 Carryng Case ....................24-0865 Instructon Manual ..................24-9388 Probe, Hose, and Water Trap / Flter Assembly ........24-7263 Quck Start Gude ..................24-9393...
  • Page 56: ServCe Centers

    United States 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Fax: 724-334-5723 Email: help@mybacharach.com Canada Bacharach of Canada, Inc. 250 Shields Court Unit #3 Markham, Ontario L3R 9W7 Canada Phone: 905-470-8985 Fax: 905-470-8963 Email: bachcan@idirect.com México Bacharach de México Playa Regatas No.
  • Page 57: Appendix A - Formulas And Tables

    Fyrite Pro Appendix A – Formulas and Tables (20.9 – O 2 max 20.9 qA = (TA – TL) x ( 20.9 – O 20.9 λ (Excess Air) = 20.9 – O Eta = (100% – qA) + [K , if TA < (T –...
  • Page 58 Fyrite Pro Notes: Instruction 24-9388...
  • Page 59 Fyrite Pro Notes: Instruction 24-9388...
  • Page 60 Headquarters: 621 Hunt Valley Circle, New Kensington, PA 15068 Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666 Website: www.mybacharach.com • E-mail: help@mybacharach.com Printed in U.S.A.