5.12.4. Starting the Data Logging Process

After turning ON data logging and setting the interval and duration, the analyzer
will pause at the LOGGING SUMMARY screen, where the currently selected
interval and duration time periods are displayed.
ENT
Press
to start the combustion test and logging process. At this time the
Combustion Test LOG screen will appear, indicating that the analyzer is now
performing a combustion test and the data is being stored in memory.
NOTE: At the top of the screen, the memory location where the
current log entry is being saved is displayed.
NOTE: If there were previously stored log entries, each new logging
session will start to store data in the next available memory location.
NOTE: At no time will new data overwrite old data. Logging will
stop when available memory is filled.

5.12.5. Ending the Data Logging Process

Data logging will stop and the pump will turn OFF after the prescribed duration,
or after all 500 memory locations are filled.
To exit the Combustion Test LOG screen, press the
display the Combustion Test HOLD screen.
NOTE: To end the logging process at any time, press the ESC
button. If the logging process was stopped by pressing ESC, it
cannot be resumed without first turning logging back ON.