5.2. Settings Menu

This menu item provides the user with the ability to configure much of the instrument
functionality to suit his specific needs. Use the ARROW DOWN button to scroll to
SETTINGS then push the MENU button to access this menu and view the listed
functions. Use the ARROW DOWN button to scroll to the function you wish to change
and press the MENU button. The functions available for modification are as follows.
PREVIOUS
SENSORS
BATTERY
KEYPAD
PREVIOUS: This menu item is used to return the user to the previous menu.
SENSORS: This menu lists only the installed sensors that provide the user with the
ability to adjust them from the menu. The choices are explained below.
TEMP:
This menu item allows the user to change the temperature units from
degrees C to degrees F or visa versa, and to adjust the temperature
value slightly to align it with another device.
available are:
Temperature adjust:
displayed temperature value slightly (+/-1 degree C or +/-2 degrees F)
to align with another device. This function can only be used once to
prevent excessive adjustment of temperature sensor without the use of
a calibration chamber. Using the ARROW DOWN button, scroll to
"ADJUST" and press the MENU button. The LCD prompts the user to
input the desired numeric value. Once the last digit has been entered,
the new temperature value will be automatically saved.
RH:
This menu function allows the user to adjust the RH value slightly to
align with another device. The available RH option is:
RH adjust: The choice is –1% or +1% RH. Use the ARROW DOWN
button to scroll to the desired choice and push the MENU button.
VOC:
The VOC (TVOCs) menu allows the user to switch the units of measure
for the PID sensor only. The VOC options available are:
Using the ARROW DOWN button, scroll down to the desired unit of
measure and push the MENU button to accept.
1509-9001 Rev 1
DEGREES C
DEGREES F
ADJUST
ADJUST
PPM
mg/m3
IEQ Chek
DISPLAY
PUMP
PASS CODE
DEFAULT
This function allows the user to tweak the
Operation Manual
The TEMP options
13