Replacing the Batteries

Main ~ Power
Reference 10 MHz
11 TO 18 VDC
EXT DC
External DC Input Connector
Figure C-2. External DC Power Socket
Replacing the Batteries
Counters equipped with the Battery option use sealed lead-acid VHS
camcorder batteries. You can obtain additional batteries of this type from
Agilent (Agilent P/N 53150-80010) and from other suppliers who carry
test equipment and/or video camera accessories.

Removing the Batteries

To remove the batteries, use the following procedure:
1
Turn off the Counter, and disconnect all external cables (including the
power cord).
2
Turn the two thumb screws that secure the battery sled to the Counter's
back panel counterclockwise until both thumb screws are completely out
of the threaded holes in the back panel (see Figure C-3).
C-4
Made in U.S.A.
with domestic and foreign content
OPTIONS
001 Oven Time Base
002 Battery
Auxillary
WARNING:
In
To avold electric shock,
or
do not remove covers.
Out
No user-serviceable parts inside.
Refer all servicing to qualified personnel.
This unit must be earth grounded.
AC POWER
100 – 130 VAC, 50/60/400 Hz 75 VA
220 – 240 VAC, 50/60 Hz 75 VA
RS-232
Main AC Power Indicator
ISM 1-A
FUSE
1.0 A T
250 V