Chapter 2 Operating Your Frequency Counter
Using Power Correction

Power Correction Example: Selecting a Correction Profile

1
Press the Shift key, and then press the Menu (Reset/Local) key.
2
Press either the up- or down-arrow key repeatedly until "PWR
CORR > OFF" is displayed as shown below:
3
Press the right-arrow key once.
The flashing pointer in the Power Correction display changes direction
to indicate that you can now use the up- and down-arrow keys to select a
power-correction profile (or select OFF to disable Power Correction).
The current setting for Power Correction (OFF or a number between 1
and 9) also flashes.
4
Press the up- and/or down-arrow key repeatedly until the number
of the power-correction profile you want to use is displayed.
For this example, choose profile number 3.
If no data points were entered into the correction profiles previously,
each of the nine power-correction profiles contains only two valid data
points—0.0 dB loss at 1 GHz and 0.0 dB loss at the highest frequency the
Counter can measure (20 GHz for the 53150A, 26.5 GHz for the 53151A,
or 46 GHz for the 53152A). The remaining eight data points in each of the
profiles initially contain values of 0.0 dB loss at 0.0 GHz. (This example
assumes that valid loss and frequency values were previously entered in
two or more of the data points in this profile.)
Operating Guide
2-33