Measuring Power (Channel 2 Only) - Agilent Technologies 53151A Operating Manual

Microwave frequency counter
Hide thumbs Also See for 53151A:
Table of Contents

Advertisement

NOTE
CAUTION
Chapter 1 Getting Started
Operating the Counter

Measuring Power (Channel 2 Only)

The Agilent 53150A/151A/152A Counter can measure signal power (in the
same frequency ranges as for frequency measurements) on Channel 2.
The example procedure for measuring power in the following diagram
assumes that the Counter has previously been set up to measure a 25 GHz
input on Channel 2.
When the power of a signal applied to the Channel 2 input exceeds the
maximum rated power for the instrument, the Counter displays HI.
The Channel 2 input path circuits contain sensitive GaAs semiconductors.
To prevent damage to these components, always adhere to standard ESD
(Electro-Static Discharge) prevention procedures, and ensure that the
maximum power specification for this channel (+27 dBm) is not exceeded.
Display
Power
1-22
Ch 2
Freq
Ch 2
Freq
dBm
Pwr
Operating Guide