Ac Current Gain Adjustment Procedure - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

Table 17
Input
Vrms
10 mV
100 mV
1 V
10V
100 V
750 V
Caution: Set the calibrator output to 0V before disconnecting from the multimeter input
terminals.

AC Current Gain Adjustment Procedure

Review the
Considerations"
1 Press
2 The primary display will show the calibration value and the secondary
3 Configure each Cal Item shown in the adjustment table below.
4 Use
5 Apply the input signal shown in the Input and Frequency columns of
Always complete tests in the same order as shown in the appropriate table.
N O T E
34405A User's and Service Guide
AC Voltage Gain Adjustment
Frequency
1kHz
1kHz
1kHZ
1kHz
1kHz
1kHz
"Test Considerations"
on page 90 sections before beginning this procedure.
ACI
to enter AC Current Gain Calibration.
display will show the reference value of the Cal Item.
(Auto) or
the table below.
Performance Tests and Calibration
Frequency as 1kHz
Function
Cal Item
AC Voltage
10 mV
100 mV
1 V
10 V
100 V
750 V
on page 67 and
(Range) to select the Cal Item.
"Gain Adjustment
4
95