Optional Capacitance Performance Verification Test

Configuration: Capacitance (CONFigure:CAPacitance)
1 Select the Capacitance function.
2 Select each range in the order shown below. Provide the indicated input
Table 13
Input Capacitance Range
[1] For the best accuracy, take a zero null measurement with open test leads, to null
out the test lead capacitance, before connecting the test leads to the calibrator.
34405A User's and Service Guide
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
Optional Capacitance Performance Verification Test
1nF
1nF
10nF
10nF
100nF
100nF
1µF
1µF
10µF
10µF
100µF
100µF
1000µF
1000µF
10000µF
10000µF
Performance Tests and Calibration
Error from Nominal
[1]
Function
1 year
Capacitance
± 28pF
± 0.15nF
± 1.5nF
± 15nF
± 0.15µF
± 1.5µF
± 15µF
± 0.25mF
4
81