Capacitance Gain Adjustment Procedure - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

6 Enter the actual applied input (see
7 Press
8 Repeat steps 3 through 7 for each gain adjustment point shown in the
9 Verify the Frequency Gain adjustments using the
Table 20
Input Voltage Frequency Function
1Vrms

Capacitance Gain Adjustment Procedure

Review the
Considerations"
1 Press
2 The primary display will show the calibration value and the secondary
3 Configure each Cal Item shown in the adjustment table below.
34405A User's and Service Guide
page 86).
Disp
to start the adjustment. The CAL annunciator flashes to
indicate the calibration is in progress.
Successful completion of each adjustment value is indicated by a
short beep and the primary display briefly showing PASS.
An adjustment failure is indicated by a long beep, the primary
display showing FAiL and a calibration error number appearing in
the secondary display. Check the input value, range, function, and
entered adjustment value to correct the problem and repeat the
adjustment step.
table.
Verification Test"
on page 76.
Frequency Gain Adjustment
1kHz
"Test Considerations"
on page 90 sections before beginning this procedure.
to enter Capacitance Gain Calibration.
display will show the reference value of Cal Item.
"Entering Adjustment Values"
Voltage as 1Vrms
Cal Item
Frequency 1kHz
on page 67 and
Performance Tests and Calibration
"Frequency Gain
"Gain Adjustment
4
on
99