Making Measurements; Measuring Ac Or Dc Voltage - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

1
Getting Started Tutorial

Making Measurements

The following pages show how to make measurement connections and how
to select measurement functions from the front panel for each of the
measurement functions.
For remote operation, refer to the MEASure Subsystem in the Agilent
34405A Online Programmer's Reference online help.

Measuring AC or DC Voltage

AC Voltage:
• Five Ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V
• Measurement Method: AC coupled true rms - measures the AC component with up to 400 VDC
• Crest Factor: Maximum 5:1 at full scale
• Input Impedance: 1 MΩ ± 2% in parallel with <100pF on all ranges
• Input Protection: 750V rms on all ranges (HI terminal)
DC Voltage:
• Five Ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 1000.00 V
• Measurement Method: Sigma Delta A-to-D converter
• Input Impedance: ~10 MΩ all ranges (typical)
• Input Protection: 1000V on all ranges (HI terminal)
ACV
DCV
20
bias on any range.
Typical ACV Display:
mV
Typical DCV Display:
mV DC
mV
AC
Range
AC
mV DC
Range
1.25A/500V FH
12A
V
Fused
HI
12A
1000VDC
rms
750VAC
AC or DC Voltage Source
LO
1.2A
500Vpk
rms
I
CAT II (300V)
34405A User's and Service Guide
+
-

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents